The effect of soft contact lens wear and time from blink on wavefront aberration measurement variation.

Sheila M Rae1, Holly C Price.   

Abstract

INTRODUCTION: Advances in contact lens design technology allow incorporation of correction of higher-order aberrations to improve or manipulate the eye's optical quality. Repeatability of aberration measurements in vivo with contact lens wear has not been established.
METHOD: Higher-order aberrations were measured using the COAS aberrometer (Wavefront Sciences, USA) in 23 participants who were free from external eye pathology. Cycloplegia with 1% Cyclopentolate HCl ensured control of accommodation and that pupil diameters exceeded analysis diameter. Tear film characteristics were assessed with the phenol red thread test and Keeler Tearscope. Variability in wavefront aberrations was assessed with and without a low powered soft contact lens. The effect of time after the blink and tear film characteristics were also investigated.
RESULTS: Standard deviations differed significantly for some fifth-order terms (p < or = 0.003) with contact lens wear. Coefficient of variation did not increase significantly with contact lens wear. Time from blink increased standard deviations in the third-order (p < or = 0.003). The effect of contact lens wear and time from blink differed between groups with different tear film characteristics.
CONCLUSIONS: Variability in measurements of aberrations increased significantly for odd-order aberrations when a contact lens was worn or between blinks. Fourth-order aberrations showed less effect of contact lens wear. Therefore, verification of the performance of lenses with aspheric optics on eye is feasible. Accurate assessment of higher-order aberrations is more difficult when a contact lens is worn. Both contact lens wear and time from blink increase variability in aberrations, which suggests a role for the tear film in increased aberrations with soft contact lens wear.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19351377     DOI: 10.1111/j.1444-0938.2009.00377.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  5 in total

1.  Spherical aberration and higher order aberrations with Balafilcon A (PureVision) and Comfilcon A (Biofinity).

Authors:  Colm McAlinden; Jonathan E Moore; Victoria E McGilligan; Tara C B Moore
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-08-06       Impact factor: 3.117

2.  Measurement of the time course of optical quality and visual deterioration during tear break-up.

Authors:  Haixia Liu; Larry Thibos; Carolyn G Begley; Arthur Bradley
Journal:  Invest Ophthalmol Vis Sci       Date:  2010-01-27       Impact factor: 4.799

3.  Objective Analysis of Pre-Lens Tear Film Stability of Daily Disposable Contact Lenses Using Ring Mire Projection.

Authors:  Sebastian Marx; Julia Eckstein; Wolfgang Sickenberger
Journal:  Clin Optom (Auckl)       Date:  2020-11-20

4.  Data quality and clinical decision-making: do we trust machines blindly?

Authors:  Konrad Pesudovs; Raymond A Applegate
Journal:  Clin Exp Optom       Date:  2009-05       Impact factor: 2.742

5.  Forward light scatter analysis of the eye in a spatially-resolved double-pass optical system.

Authors:  Jayoung Nam; Larry N Thibos; Arthur Bradley; Nikole Himebaugh; Haixia Liu
Journal:  Opt Express       Date:  2011-04-11       Impact factor: 3.894

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.