Effect of space charge on resolving power and ion loss in ion mobility spectrometry.

Adrian V Mariano1, Wansheng Su, Samar K Guharay.   

Abstract

The effect of space charge on the performance of an Ion Mobility Spectrometry (IMS) system becomes more important as the system is made smaller. We use the SIMION software package with the Statistical Diffusion Simulation (SDS) module and SIMION's new capability to solve the Poisson equation to study the effect of space charge on ion loss and resolving power in IMS systems. We consider IMS systems ranging in length from 50 mm to 150 mm and in diameter from 8.33 mm to 50 mm with a fixed electric field of 50 V/mm. We also examine a system with a length of 50 mm, a diameter of 16.7 mm, and an electric field of 16.7 V/mm. We assume that any charge density can be injected into the IMS system, and we have obtained expressions that predict the ion loss and resolving power of IMS systems as a function of input charge density and drift tube aspect ratio (length/diameter).

Entities:  

Year:  2009        PMID: 19344147     DOI: 10.1021/ac802652f

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

Review 1.  Review on ion mobility spectrometry. Part 2: hyphenated methods and effects of experimental parameters.

Authors:  R Cumeras; E Figueras; C E Davis; J I Baumbach; I Gràcia
Journal:  Analyst       Date:  2015-03-07       Impact factor: 4.616

2.  Broadscale resolving power performance of a high precision uniform field ion mobility-mass spectrometer.

Authors:  Jody C May; James N Dodds; Ruwan T Kurulugama; George C Stafford; John C Fjeldsted; John A McLean
Journal:  Analyst       Date:  2015-07-20       Impact factor: 4.616

3.  Coulombic effects in ion mobility spectrometry.

Authors:  Aleksey V Tolmachev; Brian H Clowers; Mikhail E Belov; Richard D Smith
Journal:  Anal Chem       Date:  2009-06-15       Impact factor: 6.986

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.