Literature DB >> 1934215

Electrocautery-induced pacemaker malfunction during surgery.

D Mangar1, G M Atlas, P B Kane.   

Abstract

We report a case of electrocautery-induced pacemaker failure that resulted in asystole in a 15-year-old girl scheduled for cardiac surgery. Her pacemaker was converted to the asynchronous mode the night before surgery. Electromagnetic interference from the unipolar electrocautery caused a reduction in the battery voltage, which allowed the digital circuitry, but not the voltage control oscillator (VCO), to work properly. Eventually the battery current drain caused VCO "lock-out," and pacemaker and battery failure. This report demonstrates that electrocautery-induced pacemaker failure can occur, even after conversion to asynchronous mode.

Entities:  

Mesh:

Year:  1991        PMID: 1934215     DOI: 10.1007/BF03008198

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  9 in total

1.  Anesthesia for patients with heart block and artificial cardiac pacemakers.

Authors:  J E Wynands
Journal:  Anesth Analg       Date:  1976 Sep-Oct       Impact factor: 5.108

2.  Safety precautions in the management of patients with pacemakers when electrocautery operations are performed.

Authors:  S Erdman; L Levinsky; C Servadio; E Stoupel; M J Levy
Journal:  Surg Gynecol Obstet       Date:  1988-10

3.  Suppression of a demand pacemaker by transurethral electrocautery.

Authors:  S M Lerner
Journal:  Anesth Analg       Date:  1973 Sep-Oct       Impact factor: 5.108

Review 4.  Cardiac pacemakers as an anaesthetic problem.

Authors:  D L Scott
Journal:  Anaesthesia       Date:  1970-01       Impact factor: 6.955

5.  Pacemaker malfunction from urethral electrocautery.

Authors:  R B Smith; W S Wise
Journal:  JAMA       Date:  1971-10-11       Impact factor: 56.272

6.  Electrocautery and pacemakers: management of the paced patient subject to electrocautery.

Authors:  P A Levine; G J Balady; H L Lazar; P H Belott; A J Roberts
Journal:  Ann Thorac Surg       Date:  1986-03       Impact factor: 4.330

7.  Perioperative management of the pacemaker patient.

Authors:  A B Simon
Journal:  Anesthesiology       Date:  1977-02       Impact factor: 7.892

Review 8.  Pacemakers.

Authors:  J R Zaidan
Journal:  Anesthesiology       Date:  1984-04       Impact factor: 7.892

9.  Effect of coagulating and cutting current on a demand pacemaker during transurethral resection of the prostate. A case report.

Authors:  Y K Batra; I M Bali
Journal:  Can Anaesth Soc J       Date:  1978-01
  9 in total
  9 in total

1.  Reversible prolonged pacemaker failure due to electrocautery.

Authors:  R W Peters; M R Gold
Journal:  J Interv Card Electrophysiol       Date:  1998-12       Impact factor: 1.900

Review 2.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

3.  Frequency of pacemaker malfunction associated with monopolar electrosurgery during pulse generator replacement or upgrade surgery.

Authors:  Yun Lin; Daniel P Melby; Balaji Krishnan; Selcuk Adabag; Venkatakrishna Tholakanahalli; Jian-Ming Li
Journal:  J Interv Card Electrophysiol       Date:  2017-04-17       Impact factor: 1.900

4.  Forces on cardiac implantable electronic devices during remote magnetic navigation.

Authors:  C Jilek; C Lennerz; B Stracke; H Badran; V Semmler; T Reents; S Ammar; S Fichtner; B Haller; G Hessling; I Deisenhofer; C Kolb
Journal:  Clin Res Cardiol       Date:  2012-09-29       Impact factor: 5.460

5.  Does the presence of programmable implanted devices in patients with early onset scoliosis alter typical operative and postoperative practices? A survey of spine surgeons.

Authors:  Ryan McMahon; Sara J Morgan; Jaysson T Brooks; Patrick Cahill; Ryan Fitzgerald; Ying Li; Walter H Truong
Journal:  Spine Deform       Date:  2022-02-10

6.  Ventricular fibrillation induced by coagulating mode bipolar electrocautery during pacemaker implantation in Myotonic Dystrophy type 1 patient.

Authors:  Vincenzo Russo; Anna Rago; Federica DI Meo; Nadia Della Cioppa; Andrea Antonio Papa; Maria Giovanna Russo; Gerardo Nigro
Journal:  Acta Myol       Date:  2014-12

7.  Anesthetic considerations and successful management of a patient with permanent pacemaker for cervical spine instrumentation.

Authors:  Vattipalli Sameera; Mihir P Pandia; Barkha Bindu; Keshav Goyal
Journal:  Saudi J Anaesth       Date:  2018 Oct-Dec

8.  Pacemaker Failure to Capture Caused by Electrocautery: A Rare Pacemaker Pulse Generator Change Complication.

Authors:  Aman Qureshi; Intisar Ahmed; Aamir H Khan
Journal:  Cureus       Date:  2022-08-22

9.  Implantable cardioverter defibrillator during laser transurethral resection of the prostate.

Authors:  A F Deroee; B J Cohen; J F O'Hara
Journal:  Heart Lung Vessel       Date:  2014
  9 in total