Literature DB >> 19328720

A "Buddy Wire" technique for successful OptEase filter retrieval.

Thomas Pfammatter1, Lukas Hechelhammer, Roger Pfiffner.   

Abstract

The present report describes a transfemoral, single-access OptEase filter retrieval technique that takes advantage of its particular two-level mirrored umbrella design. By threading a stiff "buddy" guide wire through the device, the filter is aligned to the long vena cava axis. This technique separates the retrieval hooklet from the adjacent venous wall, thereby facilitating its snaring. This simple technique has been successful in 10 patients after a failed standard retrieval attempt. In this series of 118 consecutively placed OptEase filters, the overall filter retrieval rate increased from 72% to 80% with this maneuver, and the rate of successful intended retrieval increased from 83% to 93%.

Entities:  

Mesh:

Year:  2009        PMID: 19328720     DOI: 10.1016/j.jvir.2009.01.024

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  1 in total

1.  Balloon-trapped Technique for Successful OptEaseFilter Retrieval.

Authors:  Katsuhiro Kondo; Hirotoshi Watanabe; Masashi Iwabuchi; Masakiyo Nobuyoshi
Journal:  Ann Vasc Dis       Date:  2011-02-17
  1 in total