Biomechanical evaluation of knotless anatomical double-layer double-row rotator cuff repair: a comparative ex vivo study.

Pierre Hepp1, Georg Osterhoff, Thomas Engel, Bastian Marquass, Thomas Klink, Christoph Josten.   

Abstract

BACKGROUND: The layered configuration of the rotator cuff tendon is not taken into account in classic rotator cuff tendon repair techniques. HYPOTHESIS: The mechanical properties of (1) the classic double-row technique, (2) a double-layer double-row (DLDR) technique in simple suture configuration, and (3) a DLDR technique in mattress suture configuration are significantly different. STUDY
DESIGN: Controlled laboratory study.
METHODS: Twenty-four sheep shoulders were assigned to 3 repair groups of full-thickness infraspinatus tears: group 1, traditional double-row repair; group 2, DLDR anchor repair with simple suture configuration; and group 3, DLDR knotless repair with mattress suture configuration. After ultrasound evaluation of the repair, each specimen was cyclically loaded with 10 to 100 N for 50 cycles. Each specimen was then loaded to failure at a rate of 1 mm/s.
RESULTS: There were no statistically significant differences among the 3 testing groups for the mean footprint area. The cyclic loading test revealed no significant difference among the 3 groups with regard to elongation. For the load-to-failure test, groups 2 and 3 showed no differences in ultimate tensile load when compared with group 1. However, when compared to group 2, group 3 was found to have significantly higher values regarding ultimate load, ultimate elongation, and energy absorbed.
CONCLUSION: The DLDR fixation techniques may provide strength of initial repair comparable with that of commonly used double-row techniques. When compared with the knotless technique with mattress sutures, simple suture configuration of DLDR repair may be too weak. Knotless DLDR rotator cuff repair may (1) restore the footprint by the use of double-row principles and (2) enable restoration of the shape and profile. CLINICAL RELEVANCE: Double-layer double-row fixation in mattress suture configuration has initial fixation strength comparable with that of the classic double-row fixation and so may potentially improve functional results of rotator cuff repair.

Entities:  

Mesh:

Year:  2009        PMID: 19307331     DOI: 10.1177/0363546509332431

Source DB:  PubMed          Journal:  Am J Sports Med        ISSN: 0363-5465            Impact factor:   6.202


  3 in total

1.  The knotless cinch-bridge technique for delaminated rotator cuff tears leads to a high healing rate and a more favorable short-term clinical outcome than suture-bridge repair.

Authors:  Philipp R Heuberer; Leo Pauzenberger; Michael S Gruber; Bernhard Kriegleder; Roman C Ostermann; Brenda Laky; Werner Anderl
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-05-06       Impact factor: 4.342

2.  High density lipoproteins and type 2 diabetes: Emerging concepts in their relationship.

Authors:  Michael S Kostapanos; Moses S Elisaf
Journal:  World J Exp Med       Date:  2014-02-20

3.  Biomechanical Comparison of Modified Suture Bridge Using Rip-Stop versus Traditional Suture Bridge for Rotator Cuff Repair.

Authors:  ZiYing Wu; Chong Zhang; Peng Zhang; TianWu Chen; ShiYi Chen; JiWu Chen
Journal:  Biomed Res Int       Date:  2016-11-15       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.