Ashwini Ojha1, Rajeshwari Rathod, Harish Padh.
Abstract
Diacerein and aceclofenac are prescribed for reducing the symptoms associated with osteoarthritis. We present a simple HPLC method with UV detection for simultaneous determination of rhein (the immediate metabolite of diacerein) and aceclofenac from human plasma samples. Sample preparation was accomplished through liquid-liquid extraction with ethyl acetate and chromatographic separation was performed on a reversed-phase ODS column. Mobile phase consisted of a mixture of acetate buffer and acetonitrile run under gradient at flow rate of 1.0 ml/min. Wavelength was set at 258 nm. The method was validated for linearity, accuracy, precision and stability. The calibration was linear over the range of 0.1-7.0 microg/ml for rhein and 0.5-20 microg/ml for aceclofenac using 500 microl plasma samples. Extraction recoveries were 85% for rhein and 70% for aceclofenac. The method can easily be adopted for high-throughput clinical and pharmacokinetic studies of above two-drug fixed dose combination formulations.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19299212 DOI: 10.1016/j.jchromb.2009.02.061