Performance evaluation of the QIAGEN EZ1 DSP Virus Kit with Abbott RealTime HIV-1, HBV and HCV assays.

George J Schneider1, Kevin G Kuper, Klara Abravaya, Carolyn R Mullen, Marion Schmidt, Astrid Bunse-Grassmann, Markus Sprenger-Haussels.   

Abstract

BACKGROUND: Automated sample preparation systems must meet the demands of routine diagnostics laboratories with regard to performance characteristics and compatibility with downstream assays.
OBJECTIVES: In this study, the performance of QIAGEN EZ1 DSP Virus Kit on the BioRobot EZ1 DSP was evaluated in combination with the Abbott RealTime HIV-1, HCV, and HBV assays, followed by thermalcycling and detection on the Abbott m2000rt platform. STUDY
DESIGN: The following performance characteristics were evaluated: linear range and precision, sensitivity, cross-contamination, effects of interfering substances and correlation.
RESULTS: Linearity was observed within the tested ranges (for HIV-1: 2.0-6.0 log copies/ml, HCV: 1.3-6.9 log IU/ml, HBV: 1.6-7.6 log copies/ml). Excellent precision was obtained (inter-assay standard deviation for HIV-1: 0.06-0.17 log copies/ml (>2.17 log copies/ml), HCV: 0.05-0.11 log IU/ml (>2.09 log IU/ml), HBV: 0.03-0.07 log copies/ml (>2.55 log copies/ml)), with good sensitivity (95% hit rates for HIV-1: 50 copies/ml, HCV: 12.5 IU/ml, HBV: 10 IU/ml). No cross-contamination was observed, as well as no negative impact of elevated levels of various interfering substances. In addition, HCV and HBV viral load measurements after BioRobot EZ1 DSP extraction correlated well with those obtained after Abbott m2000sp extraction.
CONCLUSIONS: This evaluation demonstrates that the QIAGEN EZ1 DSP Virus Kit provides an attractive solution for fully automated, low throughput sample preparation for use with the Abbott RealTime HIV-1, HCV, and HBV assays.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19286419     DOI: 10.1016/j.jcv.2009.01.003

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  2 in total

1.  Comparison of an automated nucleic acid extraction system with the column-based procedure.

Authors:  Hagen Frickmann; Rebecca Hinz; Ralf Matthias Hagen
Journal:  Eur J Microbiol Immunol (Bp)       Date:  2015-03-26

2.  Evaluation of an automated nucleic acid extractor for hepatitis C virus load quantification.

Authors:  Elisa Martró; Nerea García-Sierra; Victoria González; Verónica Saludes; Lurdes Matas; Vicenç Ausina
Journal:  J Clin Microbiol       Date:  2009-01-07       Impact factor: 5.948

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.