| Literature DB >> 19286029 |
Francisco Moragues1, Carmen Igualada.
Abstract
A new liquid chromatography-mass spectrometry with ion-trap detection method for the determination of clenbuterol, ractopamine, clenpenterol, brombuterol, hydroxyclenbuterol, mapenterol and mabuterol in animal liver and urine is described to decrease ion suppression in both matrices. The developed method consists of an aqueous extraction and a two-step clean-up. Final extract was injected in a liquid chromatograph with ion trap mass spectrometer optimized to obtain MS(3) ions which provided more specificity and better signal to noise ratio in the investigated analytes. Those steps have been essential to reach decision limits (CCalpha) values below the proposed MRPL values for each substance. The CCalpha were determined and their values were 0.05 microgkg(-1) for clenbuterol, 0.2 microgkg(-1) for ractopamine and 0.1 microgkg(-1) for the other analytes in liver and the same values in microgL(-1) for urine samples.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19286029 DOI: 10.1016/j.aca.2008.10.065
Source DB: PubMed Journal: Anal Chim Acta ISSN: 0003-2670 Impact factor: 6.558