Effect of repeated torque/mechanical loading cycles on two different abutment types in implants with internal tapered connections: an in vitro study.

Abílio Ricciardi Coppedê1, Maria da Glória Chiarello de Mattos, Renata Cristina Silveira Rodrigues, Ricardo Faria Ribeiro.   

Abstract

OBJECTIVES: Internal tapered connections were developed to improve biomechanical properties and to reduce mechanical problems found in other implant connection systems. The purpose of this study was to evaluate the effects of mechanical loading and repeated insertion/removal cycles on the torque loss of abutments with internal tapered connections.
MATERIAL AND METHODS: Sixty-eight conical implants and 68 abutments of two types were used. They were divided into four groups: groups 1 and 3 received solid abutments, and groups 2 and 4 received two-piece abutments. In groups 1 and 2, abutments were simply installed and uninstalled; torque-in and torque-out values were measured. In groups 3 and 4, abutments were installed, mechanically loaded and uninstalled; torque-in and torque-out values were measured. Under mechanical loading, two-piece abutments were frictionally locked into the implant; thus, data of group 4 were catalogued under two subgroups (4a: torque-out value necessary to loosen the fixation screw; 4b: torque-out value necessary to remove the abutment from the implant). Ten insertion/removal cycles were performed for every implant/abutment assembly. Data were analyzed with a mixed linear model (P< or =0.05).
RESULTS: Torque loss was higher in groups 4a and 2 (over 30% loss), followed by group 1 (10.5% loss), group 3 (5.4% loss) and group 4b (39% torque gain). All the results were significantly different. As the number of insertion/removal cycles increased, removal torques tended to be lower. It was concluded that mechanical loading increased removal torque of loaded abutments in comparison with unloaded abutments, and removal torque values tended to decrease as the number of insertion/removal cycles increased.

Mesh:

Substances:

Year:  2009        PMID: 19281502     DOI: 10.1111/j.1600-0501.2008.01690.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  10 in total

1.  In vitro analysis of the microbiological sealing of tapered implants after mechanical cycling.

Authors:  Deceles Cristina Costa Alves; Paulo Sérgio Perri de Carvalho; Carlos Nelson Elias; Eduardo Vedovatto; Elizabeth Ferreira Martinez
Journal:  Clin Oral Investig       Date:  2016-02-20       Impact factor: 3.573

2.  The Mechanical Behavior of a Screwless Morse Taper Implant-Abutment Connection: An In Vitro Study.

Authors:  Aimen Bagegni; Vincent Weihrauch; Kirstin Vach; Ralf Kohal
Journal:  Materials (Basel)       Date:  2022-05-08       Impact factor: 3.748

3.  Effect of cyclic loading and retightening on reverse torque value in external and internal implants.

Authors:  Woong-Rae Cho; Yoon-Hyuk Huh; Chan-Jin Park; Lee-Ra Cho
Journal:  J Adv Prosthodont       Date:  2015-08-18       Impact factor: 1.904

4.  Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

5.  Mechanical Pull-Out Test of a New Hybrid Fixture-Abutment Connection: An In Vitro Study.

Authors:  Gianmaria D'Addazio; Bruna Sinjari; Lorenzo Arcuri; Beatrice Femminella; Giovanna Murmura; Manlio Santilli; Sergio Caputi
Journal:  Materials (Basel)       Date:  2021-03-22       Impact factor: 3.623

6.  Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

Authors:  Rosario Prisco; Giuseppe Troiano; Luigi Laino; Khrystyna Zhurakivska
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

7.  Bacterial Microleakage at the Implant-Abutment Interface: An In Vitro Study.

Authors:  Simonetta D'Ercole; Tatiane Cristina Dotta; Marzieh Ramezani Farani; Niloofar Etemadi; Giovanna Iezzi; Luca Comuzzi; Adriano Piattelli; Morena Petrini
Journal:  Bioengineering (Basel)       Date:  2022-06-26

8.  Healing Abutment Distortion in Implant Prostheses: An In Vitro Study.

Authors:  Chandrashekhar Pandey; Bishwa Prakash Bhattarai; Apiwat Riddhabhaya; Natthamet Wongsirichat; Dinesh Rokaya
Journal:  J Funct Biomater       Date:  2022-06-21

9.  Bacterial infiltration and detorque at the implant abutment morse taper interface after masticatory simulation.

Authors:  Ana Paula Granja Scarabel Nogueira Bella; Alessandra Sayuri Tuzita; Ivana Barbosa Suffredini; Alberto Noriyuki Kojima; Elcio Magdalena Giovani; Alfredo Mikail Melo Mesquita
Journal:  Sci Rep       Date:  2022-10-12       Impact factor: 4.996

10.  The compressive strength of implant-abutment complex with different connection designs.

Authors:  Hsien-Ching Hung; Chiung-Shing Huang; Yu-Hwa Pan
Journal:  J Dent Sci       Date:  2019-03-28       Impact factor: 2.080

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.