Effect of stent-graft compliance on endotension after EVAR.

Willem-Maarten P F Bosman1, Jan-Willem Hinnen, Daniel J Rixen, Jaap F Hamming.   

Abstract

PURPOSE: To investigate whether, and to which extent, the type of graft has influence on the pressure in the aneurysm sac.
METHODS: The study consisted of a dynamic and a static experiment. In the dynamic experiment, a latex aneurysm was inserted into an in-vitro circulation model, with variable (80-180 mmHg) systemic systolic pressure and a pulse pressure of 40 mmHg. Seven types of devices were used to exclude the aneurysm: a latex reference graft, 3 stentless grafts made of different fabrics, and 3 different types of commercially produced stented grafts used for endovascular aneurysm repair (EVAR). The systolic and diastolic intra-aneurysm pressures (IAP) were measured, along with the pulse pressure. The mean IAP and pulse pressures were compared for each category of graft (stented/stentless) and for each graft. In the static experiments, the compliance of each graft and the latex aneurysm were determined by measuring the change in volume necessary to obtain a change in pressure. Furthermore, the IAP with each graft was measured in a nonpulsatile situation under systolic pressures varying between 60 and 140 mmHg.
RESULTS: The experiments showed that, with increasing systemic pressures, there was a small pressure (<5 mmHg) increase in the aneurysm, but there was no significant difference among grafts in the dynamic or the static experiments (p = 0.12 and p = 0.17, respectively). With the 3 stented EVAR grafts, there was less pressure transmission than with the 3 stentless grafts, but this difference was not significant (p>0.05).
CONCLUSION: This in-vitro study demonstrated that there is a clinically irrelevant pressure transmission through the graft wall and no significant difference in the pressure transmission between stented and stentless grafts.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19281281     DOI: 10.1583/08-2505.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  2 in total

Review 1.  Endotension: twenty years of a controversial term.

Authors:  Álvaro Torres-Blanco; Manuel Miralles-Hernández
Journal:  CVIR Endovasc       Date:  2021-06-05

Review 2.  A literature review of the numerical analysis of abdominal aortic aneurysms treated with endovascular stent grafts.

Authors:  David Roy; Claude Kauffmann; Sébastien Delorme; Sophie Lerouge; Guy Cloutier; Gilles Soulez
Journal:  Comput Math Methods Med       Date:  2012-09-06       Impact factor: 2.238

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.