| Literature DB >> 19280313 |
Yoshiei Shimamura1, Yoshihiko Mochizuki, Yasuyuki Yamada, Ikuko Shibasaki, Shinichiro Miyoshi.
Abstract
We encountered a case in which prosthetic mitral valve leaflets were jammed by a transmitral balloon catheter during mechanical mitral valve replacement. The catheter, which was inserted in the left ventricle to render the leaflets incompetent during de-airing, forced the leaflets into the closed position. The leaflet edges pinched the catheter lumen and obstructed balloon deflation. Excessive inflation of the balloon and extraction of the catheter before complete deflation of the balloon should be avoided to prevent a potentially serious complication.Entities:
Mesh:
Year: 2009 PMID: 19280313 DOI: 10.1007/s11748-008-0346-4
Source DB: PubMed Journal: Gen Thorac Cardiovasc Surg ISSN: 1863-6705