Literature DB >> 19277421

International normalised ratio (INR) measured on the CoaguChek S and XS compared with the laboratory for determination of precision and accuracy.

Thomas D Christensen1, Torben B Larsen, Claus Jensen, Marianne Maegaard, Benny Sørensen.   

Abstract

Oral anticoagulation therapy is monitored by the use of international normalised ratio (INR). Patients performing self-management estimate INR using a coagulometer, but studies have been partly flawed regarding the estimated precision and accuracy. The objective was to estimate the imprecision and accuracy for two different coagulometers (CoaguChek S and XS). Twenty-four patients treated with coumarin were prospectively followed for six weeks. INR's were analyzed weekly in duplicates on both coagulometers, and compared with results from the hospital laboratory. Statistical analysis included Bland-Altman plot, 95% limits of agreement, coefficient of variance (CV), and an analysis of variance using a mixed effect model. Comparing 141 duplicate measurements (a total of 564 measurements) of INR, we found that the CoaguChek S and CoaguChek XS had a precision (CV) of 3.4% and 2.3%, respectively. Regarding analytical accuracy, the INR measurements tended to be lower on the coagulometers, and regarding diagnostic accuracy the CoaguChek S and CoaguChek XS deviated more than 15% from the laboratory measurements in 40% and 43% of the measurements, respectively. In conclusion, the precision of the coagulometers was found to be good, but only the CoaguChek XS had a precision within the predefined limit of 3%. Regarding analytical accuracy, the INR measurements tended to be lower on the coagulometers, compared to the laboratory. A large proportion of measurement of the coagulometers deviated more than 15% from the laboratory measurements. Whether this will have a clinical impact awaits further studies.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19277421     DOI: 10.1160/th08-09-0601

Source DB:  PubMed          Journal:  Thromb Haemost        ISSN: 0340-6245            Impact factor:   5.249


  5 in total

Review 1.  Self-monitoring and self-management of oral anticoagulation.

Authors:  Carl J Heneghan; Josep M Garcia-Alamino; Elizabeth A Spencer; Alison M Ward; Rafael Perera; Clare Bankhead; Pablo Alonso-Coello; David Fitzmaurice; Kamal R Mahtani; Igho J Onakpoya
Journal:  Cochrane Database Syst Rev       Date:  2016-07-05

2.  Home management of INR in the public health system: feasibility of self-management of oral anticoagulation and long-term performance of individual POC devices in determining INR.

Authors:  Sabrina da Silva Saraiva; Fernanda Andrade Orsi; Marcos Pereira Santos; Tania Machado; Silmara Montalvão; Carolina Costa-Lima; Erich Vinícius de Paula; Marina Pereira Colella; Joyce Annichino-Bizzacchi
Journal:  J Thromb Thrombolysis       Date:  2016-07       Impact factor: 2.300

3.  Comparison of prothrombin time derived from CoaguChek XS and laboratory test according to fibrinogen level.

Authors:  Sue Jung Kim; Eun Young Lee; Rojin Park; Juwon Kim; Jaewoo Song
Journal:  J Clin Lab Anal       Date:  2014-03-28       Impact factor: 2.352

4.  Role of POC INR in the early stage of diagnosis of coagulopathy.

Authors:  Florian J Raimann; Marie-Louise Lindner; Christoph Martin; Lukas Jennewein; Thomas Lustenberger; Florian Piekarski; Kai Zacharowski; Christian F Weber
Journal:  Pract Lab Med       Date:  2021-05-23

Review 5.  Point-of-care testing (POCT): Current techniques and future perspectives.

Authors:  Peter B Luppa; Carolin Müller; Alice Schlichtiger; Harald Schlebusch
Journal:  Trends Analyt Chem       Date:  2011-03-21       Impact factor: 12.296