Coherent scatter computed tomography for structural and compositional stone analysis: a prospective comparison with infrared spectroscopy.

Geoffrey R Wignall1, Ian A Cunningham, John D Denstedt.   

Abstract

INTRODUCTION: Infrared spectroscopy (IRS) is a standard method of stone analysis that yields relative proportions of stone materials within a sample. IRS is destructive, as it analyzes only powdered samples, with only a fraction of the stone being analyzed. This leads to sampling bias with components over- or underestimated or even missed entirely. IRS fails to provide structural composition such as that at the stone core. Coherent scatter computed tomography (CSCT) uses diagnostic X-rays to provide detailed structural and compositional analysis of intact specimens, including detailed imaging of the stone core.
METHODS: Consecutive patients undergoing surgical treatment for stone disease were recruited for the study. Stones or fragments collected during surgery were subjected to both CSCT and IRS. The two methods were compared with respect to overall bulk composition of the stone and the ability to identify the material at the core.
RESULTS: CSCT and IRS agreed on the primary component in the majority (84.8%) of samples. CSCT detected additional components in 88.8% of stones identified as uniform by IRS. CSCT also identified a distinct stone core in 78.8% of samples, while IRS failed to detect the core component in 21.2% of these stones. In 30.3% of the stones with a core component, IRS did not identify the core mineral as the primary component.
CONCLUSION: CSCT provides superior quantitative stone analysis and is not prone to issues such as sampling error as the entire specimen is analyzed. CSCT offers excellent structural imaging of stone samples, including detailed analysis of core composition.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19265466     DOI: 10.1089/end.2008.0245

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  3 in total

1.  Iterative estimation of coherent-scattering profiles from given positions by use of a single-direction beam.

Authors:  Mitsuaki Terabe; Hiroyuki Okamoto; Kichiro Koshida
Journal:  Radiol Phys Technol       Date:  2012-05-27

2.  Analysis of urinary calculi composition by infrared spectroscopy: a prospective study of 625 patients in eastern China.

Authors:  Zhang Jing; Wang GuoZeng; Jiang Ning; Yang JiaWei; Gu Yan; Yang Fang
Journal:  Urol Res       Date:  2010-02-16

3.  Urinary stone composition analysis of 3684 patients in the eastern Shandong region of China.

Authors:  Xuebao Zhang; Jiajia Ma; Ning Wang; Chunhua Lin
Journal:  J Int Med Res       Date:  2019-11-28       Impact factor: 1.671

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.