Evaluation of platelet activation in patients supported by the Jarvik 2000* high-rotational speed impeller ventricular assist device.

Christoph Löffler1, Andreas Straub, Nicole Bassler, Katharina Pernice, Friedhelm Beyersdorf, Christoph Bode, Michael P Siegenthaler, Karlheinz Peter.   

Abstract

OBJECTIVE: The Jarvik 2000 (Jarvik Heart, Inc, New York, NY) is a thumb-sized high-speed impeller pump that is used as a ventricular assist device in patients with terminal heart failure. Because the Jarvik 2000 is designed for long-term use, it is a central question whether the mechanical forces inside the pump affect blood components. This study evaluated the potential association of the high rotational speed of the Jarvik 2000 with platelet activation, which may result in thromboembolic events.
METHODS: The study group comprised patients with terminal heart failure who were supported with the Jarvik 2000. All were men and received 100 mg aspirin daily. In 8 patients, soluble platelet activation markers (P-selectin and sCD40L), platelet counts, and hemolysis markers (haptoglobin and lactate dehydrogenase levels) were determined. In 5 patients, P-selectin expression and platelet receptor glycoprotein IIb/IIIa activation were determined with flow cytometry and compared with a control group of 5 healthy men. Platelet activation was measured at various rotational device speeds.
RESULTS: After Jarvik 2000 implantation, increased hemolysis was observed, but platelet activation markers and platelet counts were not affected. Increased rotational speed (8000 to 12,000 rpm) of the device also did not result in increased platelet activation.
CONCLUSION: The Jarvik 2000 was not associated with detectable platelet activation, despite high rotational impeller speeds.

Entities:  

Mesh:

Year:  2009        PMID: 19258099     DOI: 10.1016/j.jtcvs.2008.09.019

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  4 in total

Review 1.  Bleeding and thrombosis in chronic ventricular assist device therapy: focus on platelets.

Authors:  Antigone Koliopoulou; Stephen H McKellar; Matthew Rondina; Craig H Selzman
Journal:  Curr Opin Cardiol       Date:  2016-05       Impact factor: 2.161

2.  Clinical results with Jarvik 2000 axial flow left ventricular assist device: Osaka University Experience.

Authors:  Daisuke Yoshioka; Goro Matsumiya; Koichi Toda; Taichi Sakaguchi; Yasushi Yoshikawa; Shunsuke Saito; Hikaru Matsuda; Yoshiki Sawa
Journal:  J Artif Organs       Date:  2014-07-22       Impact factor: 1.731

3.  A quantitative comparison of mechanical blood damage parameters in rotary ventricular assist devices: shear stress, exposure time and hemolysis index.

Authors:  Katharine H Fraser; Tao Zhang; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  J Biomech Eng       Date:  2012-08       Impact factor: 2.097

4.  Comparative efficacy of in vitro and in vivo metabolized aspirin in the DeBakey ventricular assist device.

Authors:  Jawaad Sheriff; Gaurav Girdhar; Wei-Che Chiu; Jolyon Jesty; Marvin J Slepian; Danny Bluestein
Journal:  J Thromb Thrombolysis       Date:  2014-05       Impact factor: 2.300

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.