Literature DB >> 19252589

Linearity calibration of amplitude and power measurements in terahertz systems and detectors.

Mira Naftaly1, Richard Dudley.   

Abstract

We describe a device consisting of a set of silicon plates to be employed for calibrating the linearity of amplitude/power measurements of terahertz systems and detectors. The device is used to test a time-domain spectrometer, a pyroelectric sensor, and a Golay cell. The results demonstrate that the linearity of amplitude/power measurement cannot be assumed but must be experimentally verified.

Entities:  

Year:  2009        PMID: 19252589     DOI: 10.1364/ol.34.000674

Source DB:  PubMed          Journal:  Opt Lett        ISSN: 0146-9592            Impact factor:   3.776


  1 in total

1.  Measuring Open Porosity of Porous Materials Using THz-TDS and an Index-Matching Medium.

Authors:  Mira Naftaly; Iliya Tikhomirov; Peter Hou; Daniel Markl
Journal:  Sensors (Basel)       Date:  2020-05-31       Impact factor: 3.576

  1 in total