Literature DB >> 19246204

A new low-level gamma-ray spectrometry system for environmental radioactivity at the underground laboratory Felsenkeller.

M Köhler1, D Degering, M Laubenstein, P Quirin, M-O Lampert, M Hult, D Arnold, S Neumaier, J-L Reyss.   

Abstract

A low-level gamma-ray spectrometry system, based on an HPGe-detector with 92% relative efficiency recently installed in the underground laboratory Felsenkeller at 110 m water equivalent (w.e.) depth, is described. The integral background count rate normalised to the Ge-crystal mass in the energy range from 40 keV until 2.7 MeV of 0.034 s(-1)kg(-1) has been achieved by careful material selection of the detector construction material, a graded shielding construction and effective radon suppression. The detector is highly suitable for the effective surveillance of water for human consumption with decision thresholds for (226)Ra and (228)Ra in the order of some mBq L(-1).

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19246204     DOI: 10.1016/j.apradiso.2009.01.027

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Lead shielding efficiency from the gamma background measurements in the salt cavern of the Polkowice-Sieroszowice copper mine.

Authors:  Kinga Polaczek-Grelik; Jan Kisiel; Agata Walencik-Łata; Jerzy W Mietelski; Paweł Janowski; Małgorzata Harańczyk; Jan Jurkowski; Agnieszka Zalewska; Jan Kobziński; Paweł Markowski; Andrzej Sadowski
Journal:  J Radioanal Nucl Chem       Date:  2015-10-27       Impact factor: 1.371