Literature DB >> 19234913

Intra-operative combustion of Onyx embolic material.

S J Smith1, A Thomas, R D Ashpole.   

Abstract

We report a case of dangerous combustion of Tantalum/Onyx embolic agent when exposed to mono-polar diathermy during subsequent surgical resection of a partially embolised arteriovenous malformation. Surgeons should be aware that previous embolisation of a vascular malformation may make use of diathermy problematic if subsequent surgical management is required.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19234913     DOI: 10.1080/02688690802512866

Source DB:  PubMed          Journal:  Br J Neurosurg        ISSN: 0268-8697            Impact factor:   1.596


  1 in total

Review 1.  Introducing PHIL (precipitating hydrophobic injectable liquid) - a new embolic agent for the body interventional radiologist.

Authors:  Akash Prashar; Saqib Butt; Nadeem Shaida