Literature DB >> 19234830

A simple, lightweight CPAP-delivery device, composed of a three-way stopcock, for the nondependent lung.

Hiroshi Sasano1, Nobuko Sasano, Shoji Ito, Takafumi Azami, Masato Morita, Akinori Takeuchi, Kazuya Sobue.   

Abstract

PURPOSE: We aimed to introduce a simple, lightweight continuous positive airway pressure (CPAP)-delivery device for the nondependent lung during one-lung ventilation, to investigate how the type of three-way stopcocks, and the compliance and resistance of a test lung affect the relationship between the oxygen flow rate and CPAP level produced, and to examine how the device works in a clinical setting.
METHODS: In the test lung study, the bronchial blocker of a Univent tube was connected to a test lung. The effects of oxygen-flow rate, types of three-way stopcocks, and compliance and resistance of the test lung on the CPAP levels were studied. In the clinical study, the lightweight device was used to treat hypoxia in seven patients during one-lung ventilation with the bronchial blocker.
RESULTS: In the test lung study, the CPAP level produced by the device was proportional to the oxygen-flow rate, dependent on the type of three-way stopcock used, and independent of the compliance or resistance of the test lung. There was no discrepancy between the plateau pressures of the test lung and the monitoring port of an additional stopcock at any degree of compliance or resistance of the test lung at any oxygen-flow rate. Therefore, the relationship between the oxygen-flow rate and CPAP level can be ensured in advance before application to the lung, with an additional three-way stopcock of which the distal end is occluded. In the clinical study, peripheral oxygen sataration Sp(O2) improved while the CPAP level ranged from 2.8 to 5.4 cmH2O.
CONCLUSION: The lightweight CPAP delivery-device can provide variable CPAP levels by adjusting the oxygen-flow rate without real-time monitoring of the pressure.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19234830     DOI: 10.1007/s00540-008-0681-8

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  5 in total

1.  A technically simple means for administration of continuous positive airway pressure during one-lung ventilation using a Univent tube.

Authors:  V Foroughi; P E Krucylak; J Wyatt; C Bowen
Journal:  Anesth Analg       Date:  1995-09       Impact factor: 5.108

2.  Optimization of arterial oxygenation during one-lung anesthesia.

Authors:  L M Capan; H Turndorf; C Patel; S Ramanathan; A Acinapura; J Chalon
Journal:  Anesth Analg       Date:  1980-11       Impact factor: 5.108

3.  Improving oxygenation during one-lung ventilation in dogs: the effects of positive end-expiratory pressure and blood flow restriction to the nonventilated lung.

Authors:  D D Alfery; J L Benumof; F R Trousdale
Journal:  Anesthesiology       Date:  1981-10       Impact factor: 7.892

4.  Operative lung constant positive airway pressure with the Univent bronchial blocker tube.

Authors:  J L Benumof; S Gaughan; G T Ozaki
Journal:  Anesth Analg       Date:  1992-03       Impact factor: 5.108

5.  Effectiveness of low levels of nonventilated lung continuous positive airway pressure in improving arterial oxygenation during one-lung ventilation.

Authors:  C W Hogue
Journal:  Anesth Analg       Date:  1994-08       Impact factor: 5.108

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.