Literature DB >> 19234017

Invasive arterial blood pressure monitoring in an out-of-hospital setting: an observational study.

J Sende1, P Jabre, B Leroux, C Penet, E Lecarpentier, M Khalid, A Margenet, J Marty, X Combes.   

Abstract

BACKGROUND: Non-invasive arterial blood pressure measurement is often inaccurate in emergency unstable patients. A study was undertaken to assess the feasibility of out-of-hospital intra-arterial catheterisation in haemodynamically unstable patients and to evaluate the correlation between invasive and non-invasive arterial pressure values.
METHODS: In this prospective 2-year observational study conducted by mobile emergency medical units, the success rate of arterial catheterisation was calculated and blood pressure values measured invasively and non-invasively after successful catheterisation were compared.
RESULTS: 94 patients were included. The success rate for catheterisation (44 radial access, 50 femoral access) was 86% (95% CI 79% to 93%). Bias and precision in invasive versus non-invasive comparisons were -0.1, 38 mm Hg for systolic pressure and 4.2, 27 mm Hg for diastolic pressure. Values differed by more than 20 mm Hg in over 40% of patients. Invasive measurement led to 79 changes in vasoactive treatment in 51 patients.
CONCLUSION: Emergency out-of-hospital invasive arterial blood pressure monitoring in haemodynamically unstable patients is highly feasible. Discrepancies between invasive and non-invasive measurements are common and highlight the value of early out-of-hospital monitoring.

Entities:  

Mesh:

Year:  2009        PMID: 19234017     DOI: 10.1136/emj.2008.060608

Source DB:  PubMed          Journal:  Emerg Med J        ISSN: 1472-0205            Impact factor:   2.740


  1 in total

1.  A non-invasive method to assess cerebral perfusion pressure in geriatric patients with suspected cerebrovascular disease.

Authors:  Bo Liu; Qi Li; Kewei Li; Nan Deng; Peng He; Chunchang Qin; Deyu Yang; Zhiwei Li; Peng Xie
Journal:  PLoS One       Date:  2015-03-19       Impact factor: 3.240