| Literature DB >> 1921876 |
Abstract
The Batho power-law formula is in common use in many treatment planning systems to correct for the presence of lungs and other inhomogeneities. While giving excellent agreement with measurement for Cobalt-60 radiation, it tends to underestimate the lung correction required for higher energy x rays and is undefined for distances beyond an interface less than the buildup distance. This paper suggests a simple modification that greatly improves the agreement with measured data and gives a continuously defined function at all depths. Measurements have been made in a polystyrene and cork phantom to simulate the effects of lung; data are presented for beams of 8-MV x rays, 16-MV x rays, and Cobalt-60 gamma rays.Entities:
Mesh:
Substances:
Year: 1991 PMID: 1921876 DOI: 10.1118/1.596665
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071