Initial radiation experience evaluating early tolerance and toxicities in patients undergoing accelerated partial breast irradiation using the Contura Multi-Lumen Balloon breast brachytherapy catheter.

Sheree Brown1, Mark McLaughlin2, Keith Pope1, Kenneth Haile1, Laurie Hughes1, Philip Z Israel3.   

Abstract

PURPOSE: We reviewed our institution's experience treating patients with the Contura Multi-Lumen Balloon (SenoRx, Inc., Irvine, CA) breast brachytherapy catheter to deliver accelerated partial breast irradiation. METHODS AND MATERIALS: Forty-one patients treated with breast-conserving therapy received adjuvant radiation using the Contura catheter (34Gy in 3.4Gy fractions). Thirteen patients had Stage 0, 21 had Stage I, and 7 had Stage II breast cancer. Median followup was 8 months (range, 1-17).
RESULTS: Median, minimum skin spacing was 10mm (range, 2-17). Median, maximum skin doses (% of prescribed dose [PD]) were 99.7 (range, 57.1-124.1). Eight patients were treated with a skin spacing <or=5mm and 2 had a spacing of 2mm. Median, maximum rib doses were 102.6% of PD (10.0-187.7), and the median percentage of the planning target volume for evaluation (PTV_EVAL) receiving 95% of the PD was 98.8 (range, 79.4-107.4). The median volume receiving 200% of the PD was 5.7cc (range, 1.3-9.9). The percentage of patients with excellent/good cosmetic results at 6 months (n=15) and 12 months (n=12) was 100%. Patient tolerance was assessed on a scale 0-10 (0=no pain, 10=requiring narcotic analgesics). In 37 out of 38 patients, pain was graded <or=3 at the time of catheter insertion. Four breast infections (11%) and one transient symptomatic seroma (3%) developed.
CONCLUSION: Adjuvant accelerated partial breast irradiation using the Contura Multi-Lumen Balloon catheter exhibited similar toxicities to standard single lumen balloon brachytherapy with improvements in dosimetric capabilities (i.e., reduced skin and rib doses and improved PTV_EVAL coverage).

Entities:  

Mesh:

Year:  2009        PMID: 19217353     DOI: 10.1016/j.brachy.2008.11.013

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  10 in total

Review 1.  Recent developments and best practice in brachytherapy treatment planning.

Authors:  C D Lee
Journal:  Br J Radiol       Date:  2014-06-02       Impact factor: 3.039

2.  Choice of management of southern Chinese BRCA mutation carriers.

Authors:  Ava Kwong; Connie H N Wong; Catherine Shea; Dacita T K Suen; Catherine L Y Choi
Journal:  World J Surg       Date:  2010-07       Impact factor: 3.352

3.  Radiation therapy in early-stage invasive breast cancer.

Authors:  Ray Lin; Prabhakar Tripuraneni
Journal:  Indian J Surg Oncol       Date:  2011-05-06

Review 4.  Accelerated Partial Breast Irradiation (APBI): A review of available techniques.

Authors:  Christopher F Njeh; Mark W Saunders; Christian M Langton
Journal:  Radiat Oncol       Date:  2010-10-04       Impact factor: 3.481

5.  Attitudes and compliance of clinical management after genetic testing for hereditary breast and ovarian cancer among high-risk Southern Chinese females with breast cancer history.

Authors:  Ava Kwong; Annie Tsz-Wai Chu; Christine Teen-Sum Wu; Desiree Man-Sik Tse
Journal:  Fam Cancer       Date:  2014-09       Impact factor: 2.375

6.  Transforming growth factor β-1 (TGF-β1) is a serum biomarker of radiation induced fibrosis in patients treated with intracavitary accelerated partial breast irradiation: preliminary results of a prospective study.

Authors:  Dustin L Boothe; Shana Coplowitz; Eleni Greenwood; Christian L Barney; Paul J Christos; Bhupesh Parashar; Dattatreyudu Nori; K S Clifford Chao; A Gabriella Wernicke
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-10-16       Impact factor: 7.038

7.  Dosimetric Improvements in Balloon Based Brachytherapy Using the Contura® Multi-Lumen Balloon (MLB) Catheter to Deliver Accelerated Partial Breast Irradiation.

Authors:  Frank A Vicini; Arthur Douglas; Dorin Todor; Thomas Julian; Maureen Lyden
Journal:  J Contemp Brachytherapy       Date:  2010-04-01

8.  Acute toxicity and early cosmetic outcome in patients treated with multilumen balloon brachytherapy with skin spacing ≤ 7.0 millimeters.

Authors:  Kiwhoon Lee; Amy Quillo; Dustin Dillon; Keith Sowards; Allison Hatmaker; Anthony Dragun
Journal:  J Contemp Brachytherapy       Date:  2012-03-30

9.  Comparative dosimetric findings using accelerated partial breast irradiation across five catheter subtypes.

Authors:  Zaker Rana; Nadim M Nasr; Huaying Ji; Virginia Lorio; Stephanie Akbari; Molly Sebastian; Mami Martin; Robert L Hong
Journal:  Radiat Oncol       Date:  2015-07-31       Impact factor: 3.481

10.  Dosimetric evaluation of multilumen intracavitary balloon applicator rotation in high-dose-rate brachytherapy for breast cancer.

Authors:  Yongbok Kim; Mark G Trombetta
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.