Falsely elevated whole blood tacrolimus concentrations due to interference in an affinity column-mediated immunoassay method on Xpand dimension.

Caroline Barau1, Carlos Frangie, Cécile Goujard, Olivier Tribut, François Parant, Anne-Marie Taburet, Antoine Durrbach, Valérie Furlan.   

Abstract

The whole blood concentration of tacrolimus is required for therapeutic drug monitoring of this immunosuppressive drug. Abnormal tacrolimus levels affect its efficacy or toxicity, leading to changes in its dosage. Here, we report analytical interference in the affinity column-mediated immunoassay tacrolimus method on the Xpand autoanalyzer in a kidney transplant human immunodeficiency virus-infected patient. Tacrolimus concentrations obtained by affinity column-mediated immunoassay are 3- to 7-fold higher than measurements with the enzyme multiplied immunoassay technique assay. The cause of this interference remains unknown. However, it would be necessary to identify this type of interference to measure tacrolimus concentration with another method to avoid analytical error, which may lead to a poor clinical outcome.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19214146     DOI: 10.1097/FTD.0b013e31819a7c19

Source DB:  PubMed          Journal:  Ther Drug Monit        ISSN: 0163-4356            Impact factor:   3.681


  1 in total

1.  False Elevation of the Blood Tacrolimus Concentration, as Assessed by an Affinity Column-mediated Immunoassay (ACMIA), Led to Acute T Cell-mediated Rejection after Kidney Transplantation.

Authors:  Momoko Kono; Jumpei Hasegawa; Hina Ogawa; Kanae Yoshikawa; Ayumi Ishiwatari; Sachiko Wakai; Kazunari Tanabe; Hiroki Shirakawa
Journal:  Intern Med       Date:  2017-12-27       Impact factor: 1.271