Jonathan P Williams1, Hannah Jackson, Brian N Green.
Abstract
When analyzed by routine cation exchange high performance liquid chromatography (HPLC), the Hb A(1c) peak of a Caucasian diabetic patient from Yorkshire, UK, appeared to be an incompletely resolved doublet. One component (5.5%) eluted at the normal time for Hb A(1c), whereas the other component (6.6%) eluted slightly later. The HPLC trace was otherwise normal. Analysis of the diabetic patient's blood and a tryptic digest thereof by electrospray ionization mass spectrometry (ESI-MS) identified the Hb Belleville trait. To relate Hb A(1c) determined by HPLC to alpha and beta chain glycation determined by ESI-MS, several normal blood samples (5-14% Hb A(1c)) were analyzed by both techniques. The Hb A(1c) levels derived from the alpha and beta chain glycation levels of the diabetic patient (12.9 and 12.4%, respectively) agreed with the sum of the two peaks (12.1%) in the HPLC trace. Similarly, Hb Belleville was detected and identified in another Caucasian, also from Yorkshire, with normal Hb A(1c).Entities: