Although total elbow arthroplasty is a reliable treatment method for the arthritic joint, revision is necessary if loosening, instability, wear, fracture or infection occurs. We report 12 patients with 19 months (10-29) follow-up after revision arthroplasty for painful loosening in six cases, periprosthetic fracture in three, gross instability in one and skin perforation with possible infection in two. All cases presented with severe metallosis. Fractures were seen in four, triceps insufficiency in three and skin perforation in two. All patients were treated with exchange arthroplasty using the Coonrad-Morrey prosthesis. Three patients had in addition strut-allografting. The two patients with skin perforation and possible infection had a staged procedure. Mayo Elbow Performance Score improved from 24 to 87 with uneventful healing, fracture consolidation and allograft incorporation. At follow-up, triceps insufficiency persisted in all three cases. Transient neuropathy was present in five patients (4 ulnar, 1 radial). Metallosis in total elbow arthroplasty may be associated with severe tissue damage. A staging method is presented.

Entities:  

Mesh:

Year:  2008        PMID: 19205321

Source DB:  PubMed          Journal:  Acta Orthop Belg        ISSN: 0001-6462            Impact factor:   0.500


  6 in total

1.  Metallosis following silicone metacarpophalangeal joint arthroplasties with grommets: case report.

Authors:  Imran K Choudhry; Joyce M Wilson; Peter J Stern
Journal:  Hand (N Y)       Date:  2012-06

Review 2.  What design and material factors impact the wear and corrosion performance in total elbow arthroplasties?

Authors:  Mark P Figgie; Timothy M Wright; Denise Drinkwater
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

3.  Metallosis and Metal-Induced Synovitis Following Total Knee Arthroplasty: Review of Radiographic and CT Findings.

Authors:  Jason W Romesburg; Paul L Wasserman; Candace H Schoppe
Journal:  J Radiol Case Rep       Date:  2010-09-01

4.  Severe metallosis-related osteolysis as a cause of failure after total knee replacement.

Authors:  Khaled Hamed Salem; Norbert Lindner; Markus Tingart; Alyaa Diaa Elmoghazy
Journal:  J Clin Orthop Trauma       Date:  2019-04-18

5.  Pseudotumor due to metallosis after total elbow arthroplasty.

Authors:  Roberto Rotini; Graziano Bettelli; Michele Cavaciocchi; Lucia Savarino
Journal:  Indian J Orthop       Date:  2017 Jan-Feb       Impact factor: 1.251

Review 6.  Silicone and Pyrocarbon Artificial Finger Joints.

Authors:  F A Alnaimat; H A Owida; A Al Sharah; M Alhaj; Mohammad Hassan
Journal:  Appl Bionics Biomech       Date:  2021-06-03       Impact factor: 1.781

  6 in total