Sensor fusion using a hybrid median filter for artifact removal in intraoperative heart rate monitoring.

Ping Yang1, Guy A Dumont, J Mark Ansermino.   

Abstract

OBJECTIVE: Intraoperative heart rate is routinely measured independently from the ECG monitor, pulse oximeter, and the invasive blood pressure monitor if available. The presence of artifacts, in one or more of theses signals, especially sustained artifacts, represents a critical challenge for physiological monitoring. When temporal filters are used to suppress sustained artifacts, unwanted delays or signal distortion are often introduced. The aim of this study was to remove artifacts and derive accurate estimates for the heart rate signal by using measurement redundancy.
METHOD: Heart rate measurements from multiple sensors and previous estimates that fall in a short moving window were treated as samples of the same heart rate. A hybrid median filter was used to align these samples into one ordinal series and to select the median as the fused estimate. This method can successfully remove artifacts that are sustained for shorter than half the length of the filter window, or artifacts that are sustained for a longer duration but presented in no more than half of the sensors.
RESULTS: The method was tested on both simulated and clinical cases. The performance of the hybrid median filter in the simulated study was compared with that of a two-step estimation process, comprising a threshold-controlled artifact-removal module and a Kalman filter. The estimation accuracy of the hybrid median filter is better than that of the Kalman filter in the presence of artifacts.
CONCLUSIONS: The hybrid median filter combines the structural and temporal information from two or more sensors and generates a robust estimate of heart rate without requiring strict assumptions about the signal's characteristics. This method is intuitive, computationally simple, and the performance can be easily adjusted. These considerable benefits make this method highly suitable for clinical use.

Mesh:

Year:  2009        PMID: 19199059     DOI: 10.1007/s10877-009-9163-2

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  9 in total

1.  A new method for pulse oximetry possessing inherent insensitivity to artifact.

Authors:  M J Hayes; P R Smith
Journal:  IEEE Trans Biomed Eng       Date:  2001-04       Impact factor: 4.538

Review 2.  Monitoring arterial blood pressure: what you may not know.

Authors:  Beate H McGhee; Elizabeth J Bridges
Journal:  Crit Care Nurse       Date:  2002-04       Impact factor: 1.708

3.  On-line novelty detection for artefact identification in automatic anaesthesia record keeping.

Authors:  Stephen W Hoare; David Asbridge; Paul C W Beatty
Journal:  Med Eng Phys       Date:  2002-12       Impact factor: 2.242

4.  Integrated monitoring and analysis for early warning of patient deterioration.

Authors:  F Ismail; M Davies
Journal:  Br J Anaesth       Date:  2007-01       Impact factor: 9.166

5.  Adaptive change detection in heart rate trend monitoring in anesthetized children.

Authors:  Ping Yang; Guy Dumont; J Mark Ansermino
Journal:  IEEE Trans Biomed Eng       Date:  2006-11       Impact factor: 4.538

Review 6.  The problem of artifacts in patient monitor data during surgery: a clinical and methodological review.

Authors:  George Takla; John H Petre; D John Doyle; Mayumi Horibe; Bala Gopakumaran
Journal:  Anesth Analg       Date:  2006-11       Impact factor: 5.108

7.  A robust sensor fusion method for heart rate estimation.

Authors:  M H Ebrahim; J M Feldman; I Bar-Kana
Journal:  J Clin Monit       Date:  1997-11

8.  Robust sensor fusion improves heart rate estimation: clinical evaluation.

Authors:  J M Feldman; M H Ebrahim; I Bar-Kana
Journal:  J Clin Monit       Date:  1997-11

9.  Monitoring renal transplants: an application of the multiprocess Kalman filter.

Authors:  A F Smith; M West
Journal:  Biometrics       Date:  1983-12       Impact factor: 2.571

  9 in total
  4 in total

1.  Sensor fusion methods for reducing false alarms in heart rate monitoring.

Authors:  Gabriel Borges; Valner Brusamarello
Journal:  J Clin Monit Comput       Date:  2015-10-06       Impact factor: 2.502

Review 2.  Decision fusion in healthcare and medicine: a narrative review.

Authors:  Elham Nazari; Rizwana Biviji; Danial Roshandel; Reza Pour; Mohammad Hasan Shahriari; Amin Mehrabian; Hamed Tabesh
Journal:  Mhealth       Date:  2022-01-20

3.  Machine Learning and Decision Support in Critical Care.

Authors:  Alistair E W Johnson; Mohammad M Ghassemi; Shamim Nemati; Katherine E Niehaus; David A Clifton; Gari D Clifford
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2016-01-25       Impact factor: 10.961

4.  Estimating respiratory and heart rates from the correntropy spectral density of the photoplethysmogram.

Authors:  Ainara Garde; Walter Karlen; J Mark Ansermino; Guy A Dumont
Journal:  PLoS One       Date:  2014-01-22       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.