Reliability of infrared ear thermometry in the prediction of rectal temperature in older inpatients.

Simon Smitz1, Ann Van de Winckel, Marc-François Smitz.   

Abstract

AIMS AND
OBJECTIVES: To assess the capability of infrared ear thermometry accurately to predict rectal temperature in older patients.
BACKGROUND: Infrared ear thermometry is now commonly used for predicting body temperature in older patients. However, ear thermometry has been insufficiently evaluated in geriatric patients.
DESIGN: Prospective, convenience sample, unblinded study.
METHODS: All patients (or their guardians) gave informed consent. Patients hospitalised in a geriatric unit underwent sequential ear and rectal temperatures measurements using two different models of infrared ear thermometers (ThermoScan and Genius) and a rectal probe, respectively. After a brief otoscopic examination, ear temperatures were measured twice at both ears with each thermometer, the highest of four measurements being retained for analysis. The rectal temperature was the reference standard.
RESULTS: Hundred patients (31 males), aged 81 (SD 7) years completed the study. The mean rectal temperature was 37.3 degrees C (SD 0.7) degrees C (range 36.3-40.7 degrees C). Eighteen patients were febrile (rectal temperature >or= 37.8 degrees C). The mean bias between rectal and ear temperatures as measured with the ThermoScan was -0.20 degrees C (SD 0.32) degrees C and the 95% limits of agreement were -0.83 degrees C and 0.42 degrees C (95% CI, -0.88-0.48 degrees C). Using the Genius, the corresponding figures were -0.56 degrees C (SD 0.39) degrees C, -1.32 degrees C and 0.20 degrees C (95% CI, -1.39-0.27 degrees C). After correction for bias, the ThermoScan predicted the level of fever with a maximum error of 0.7 degrees C (mean error 0.3 degrees C). Using the Genius, the maximum error and the mean error were 1.6 degrees C and 0.4 degrees C, respectively.
CONCLUSIONS: Infrared ear thermometry can predict rectal temperature in normothermic and in febrile inpatients with an acceptable level of accuracy. However, the predictive accuracy depends on both operator technique and quality of instrumentation. RELEVANCE TO CLINICAL PRACTICE: Proper technique (measuring in both ears) and optimal instrumentation (model of ear thermometer) are essential for accuracy.

Entities:  

Mesh:

Year:  2009        PMID: 19191993     DOI: 10.1111/j.1365-2702.2008.02565.x

Source DB:  PubMed          Journal:  J Clin Nurs        ISSN: 0962-1067            Impact factor:   3.036


  4 in total

1.  Urine temperature as an index for the core temperature of industrial workers in hot or cold environments.

Authors:  Shoko Kawanami; Seichi Horie; Jinro Inoue; Makiko Yamashita
Journal:  Int J Biometeorol       Date:  2012-01-26       Impact factor: 3.787

2.  A novel method for measuring sublingual temperature using conventional non-contact forehead thermometer.

Authors:  Chakrapani Mahabala; Pradeepa H Dakappa; Arjun R Gupta
Journal:  F1000Res       Date:  2022-01-07

3.  Intra-rater reliability when using a tympanic thermometer under different self-measurement conditions.

Authors:  Won-Gyu Yoo
Journal:  J Phys Ther Sci       Date:  2016-07-29

4.  Longitudinal assessment of kidney function in migrant farm workers.

Authors:  Nicolás López-Gálvez; Rietta Wagoner; Robert A Canales; Kacey Ernst; Jefferey L Burgess; Jill de Zapien; Cecilia Rosales; Paloma Beamer
Journal:  Environ Res       Date:  2021-07-14       Impact factor: 6.498

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.