Literature DB >> 19191212

Unrecognized hypoglycemia due to maltodextrin interference with bedside glucometry.

Barbara M Kirrane1, Elizabeth A Duthie, Lewis S Nelson.   

Abstract

INTRODUCTION: Glucometry is widely used to confirm or exclude hypoglycemia in patients with suggestive clinical findings. Nonglucose sugars may be detected by certain types of glucometers, causing false elevation of the glucometer analysis of the blood sugar. Since these other sugars are not functionally glucose and may even induce excess insulin release, clinical hypoglycemia may be missed. CASE REPORT: We report a 79-year-old man on enteral feeds containing maltodextrin, a glucose polymer, who had persistently high glucometer-measured blood glucose despite normal blood glucose measured by formal laboratory analysis. DISCUSSION: Excess insulin administration, based on the erroneous glucometer reading, may have caused unrecognized fatal clinical hypoglycemia. This has been reported following intravenous administration of related nonglucose sugars but not with enteral maltodextrin. Further study is required to confirm the effects of maltodextrin on glucometry.
CONCLUSION: False elevation of blood glucose measured on certain point-of-care glucometers can occur following the oral administration of maltodextrin.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19191212      PMCID: PMC3550341          DOI: 10.1007/BF03160976

Source DB:  PubMed          Journal:  J Med Toxicol        ISSN: 1556-9039


  10 in total

1.  Monitoring neonatal hypoglycemia with the Accu-chek advantage II glucose meter: the cautionary tale of galactosemia.

Authors:  Julie D Newman; C Andrew Ramsden; Nicholas D H Balazs
Journal:  Clin Chem       Date:  2002-11       Impact factor: 8.327

2.  Spurious hyperglycaemia and icodextrin in peritoneal dialysis fluid.

Authors:  Stephen G Riley; James Chess; Kieron L Donovan; John D Williams
Journal:  BMJ       Date:  2003-09-13

3.  Glucose monitoring for diabetic patients using icodextrin.

Authors:  Run Wang; Line Skoufos; Leo Martis
Journal:  Perit Dial Int       Date:  2004 May-Jun       Impact factor: 1.756

4.  A quantitative appraisal of interference by icodextrin metabolites in point-of-care glucose analyses.

Authors:  Joke J Apperloo; Huib L Vader
Journal:  Clin Chem Lab Med       Date:  2005       Impact factor: 3.694

5.  Positive interference of icodextrin metabolites in some enzymatic glucose methods.

Authors:  W Janssen; G Harff; M Caers; A Schellekens
Journal:  Clin Chem       Date:  1998-11       Impact factor: 8.327

6.  Important causes of hypoglycaemia in patients with diabetes on peritoneal dialysis.

Authors:  S Mehmet; G Quan; S Thomas; D Goldsmith
Journal:  Diabet Med       Date:  2001-08       Impact factor: 4.359

7.  Atrophic gastritis is associated with increased sucrose permeability related to chronic inflammation.

Authors:  Camilla Sjöstedt Zsigmond; Ulf Hannestad; Lennart Franzén; Johan D Söderholm; Kurt Borch
Journal:  Digestion       Date:  2005-10-20       Impact factor: 3.216

8.  Xylitol, a partial homologue ot alpha-D-glucopyranose: potent stimulator of insulin release in dogs.

Authors:  T Asano; B Z Greenberg; R V Wittmers; F C Goetz
Journal:  Endocrinology       Date:  1977-02       Impact factor: 4.736

9.  Plasma-equivalent glucose at the point-of-care: evaluation of Roche Accu-Chek Inform and Abbott Precision PCx glucose meters.

Authors:  Timothy Ghys; Wim Goedhuys; Katrien Spincemaille; Frans Gorus; Erik Gerlo
Journal:  Clin Chim Acta       Date:  2007-08-07       Impact factor: 3.786

10.  A simple, non-invasive marker of gastric damage: sucrose permeability.

Authors:  L R Sutherland; M Verhoef; J L Wallace; G Van Rosendaal; R Crutcher; J B Meddings
Journal:  Lancet       Date:  1994-04-23       Impact factor: 79.321

  10 in total
  3 in total

1.  White fat, factitious hyperglycemia, and the role of FDG PET to enhance understanding of adipocyte metabolism.

Authors:  Michael S Hofman; Rodney J Hicks
Journal:  EJNMMI Res       Date:  2011-06-07       Impact factor: 3.138

2.  Pseudohyperglycaemia in a comatose patient after picking cherries.

Authors:  Clément Derkenne; Antoine Lamblin; Daniel Jost; Jean-Pierre Tourtier
Journal:  BMJ Case Rep       Date:  2016-10-28

3.  A disposable tear glucose biosensor--part 3: assessment of enzymatic specificity.

Authors:  Kenneth Lan; Kenyon McAferty; Pankti Shah; Erica Lieberman; Dharmendra R Patel; Curtiss B Cook; Jeffrey T La Belle
Journal:  J Diabetes Sci Technol       Date:  2011-09-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.