Improvement in magnetic field immunity of externally-coupled transcutaneous energy transmission system for a totally implantable artificial heart.

Takahiko Yamamoto1, Kohji Koshiji, Akihiko Homma, Eisuke Tatsumi, Yoshiyuki Taenaka.   

Abstract

Transcutaneous energy transmission (TET) that uses electromagnetic induction between the external and internal coils of a transformer is the most promising method to supply driving energy to a totally implantable artificial heart without invasion. Induction-heating (IH) cookers generate magnetic flux, and if a cooker is operated near a transcutaneous transformer, the magnetic flux generated will link with the external and internal coils of the transcutaneous transformer. This will affect the performance of the TET and the artificial heart system. Hence, it is necessary to improve the magnetic field immunity of the TET system. During operation of the system, if the transcutaneous transformer is in close proximity to an IH cooker, the electric power generated by the cooker and coupled to the transformer can drive the artificial heart system. To prevent this coupling, the external coil was shielded with a conductive shield that had a slit in it. This reduces the coupling between the transformer and the magnetic field generated by the induction cooker. However, the temperature of the shield increased due to heating by eddy currents. The temperature of the shield can be reduced by separating the IH cooker and the shield.

Entities:  

Mesh:

Year:  2008        PMID: 19184291     DOI: 10.1007/s10047-008-0435-2

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  2 in total

Review 1.  Journal of Artificial Organs 2008: the year in review.

Authors:  Y Sawa; E Tatsumi; A Funakubo; T Horiuchi; K Iwasaki; A Kishida; T Masuzawa; K Matsuda; M Nishimura; T Nishimura; Y Tomizawa; T Yamaoka; H Watanabe
Journal:  J Artif Organs       Date:  2009-03-29       Impact factor: 1.731

2.  Primary side control of load voltage for transcutaneous energy transmission.

Authors:  Yang Fu; Liang Hu; Xiaodong Ruan; Xin Fu
Journal:  J Artif Organs       Date:  2015-09-08       Impact factor: 1.731

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.