Literature DB >> 19181973

Patient and surgeon radiation exposure: comparison of standard and mini-C-arm fluoroscopy.

Brian D Giordano1, Judith F Baumhauer, Thomas L Morgan, Glenn R Rechtine.   

Abstract

BACKGROUND: Use of c-arm fluoroscopy is common in the operating room, outpatient clinic, and emergency department. Consequently, there is a concern regarding radiation exposure. Mini-c-arm fluoroscopes have gained popularity; however, few studies have quantified exposure during mini-c-arm imaging of a body part larger than a hand or wrist. The purpose of this study was to measure radiation exposure sustained by the patient and surgeon during the use of large and mini-c-arm fluoroscopy of an ankle specimen.
METHODS: Standard and mini-c-arm fluoroscopes were used to image a cadaver ankle specimen, which was suspended on an adjustable platform. Dosimeters were mounted at specific positions and angulations to detect direct and scatter radiation. Testing was conducted under various scenarios that altered the proximity of the specimen and the radiation source. We attempted to capture a range of exposure data under conditions ranging from a best to a worst-case scenario, as one may encounter in a procedural setting.
RESULTS: With all configurations tested, measurable exposure during use of the large-c-arm fluoroscope was considerably higher than that during use of the mini-c-arm fluoroscope. Patient and surgeon exposure was notably amplified when the specimen was positioned closer to the x-ray source. The exposure values that we measured during ankle fluoroscopy were consistently higher than the exposure values that have been recorded previously during hand or wrist imaging.
CONCLUSIONS: Exposure of the patient and surgeon to radiation depends on the tissue density and the shape of the imaged extremity. Elevated exposure levels can be expected when larger body parts are imaged or when the extremity is positioned closer to the x-ray source. When it is possible to satisfactorily image an extremity with use of the mini c-arm, it should be chosen over its larger counterpart.

Entities:  

Mesh:

Year:  2009        PMID: 19181973     DOI: 10.2106/JBJS.H.00407

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  14 in total

1.  Radiation risk from fluoroscopically-assisted anterior cruciate ligament reconstruction.

Authors:  J P Chitnavis; A Karthikesalingam; A Karthikesaligam; A Macdonald; C Brown
Journal:  Ann R Coll Surg Engl       Date:  2010-05       Impact factor: 1.891

2.  A simple technique to reduce surgeon radiation exposure during the use of the mini C-arm fluoroscope.

Authors:  J Smith; R Frood; S Southern
Journal:  Ann R Coll Surg Engl       Date:  2015-03       Impact factor: 1.891

Review 3.  Radiation protection and standardization.

Authors:  O P Lakhwani; Vipin Dalal; Mohit Jindal; Ashok Nagala
Journal:  J Clin Orthop Trauma       Date:  2018-08-07

Review 4.  The use of ionising radiation in orthopaedic surgery: principles, regulations and managing risk to surgeons and patients.

Authors:  Mohsen Raza; James Houston; Ryan Geleit; Rachel Williams; Alex Trompeter
Journal:  Eur J Orthop Surg Traumatol       Date:  2021-04-07

5.  Mini C-Arm Fluoroscopy: Does Its Configuration Matter for Radiation Exposure to the Surgeon?

Authors:  Talia Chapman; Dennis P Martin; Christopher Williamson; Brian Tinsley; Mark L Wang; Asif M Ilyas
Journal:  Hand (N Y)       Date:  2017-06-23

Review 6.  C-arm fluoroscopy in orthopaedic surgical practice.

Authors:  Ishaq Ojodu; Ayodele Ogunsemoyin; Sascha Hopp; Tim Pohlemann; Oluwole Ige; Oluwaseun Akinola
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-05-23

7.  Volume-of-change cone-beam CT for image-guided surgery.

Authors:  Junghoon Lee; J Webster Stayman; Yoshito Otake; Sebastian Schafer; Wojciech Zbijewski; A Jay Khanna; Jerry L Prince; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2012-07-17       Impact factor: 3.609

8.  Radiation Exposure and Hand Dominance Using Mini C-Arm Fluoroscopy in Hand Surgery.

Authors:  Corey B Fuller; Montri D Wongworawat; Barth B Riedel
Journal:  Hand (N Y)       Date:  2016-02-02

9.  A Radiation Safety Training Program Results in Reduced Radiation Exposure for Orthopaedic Residents Using the Mini C-arm.

Authors:  David Gendelberg; William Hennrikus; Jennifer Slough; Douglas Armstrong; Steven King
Journal:  Clin Orthop Relat Res       Date:  2015-11-13       Impact factor: 4.176