Literature DB >> 19180678

Coupling the Vibrant Soundbridge to cochlea round window: auditory results in patients with mixed hearing loss.

Achille M Beltrame1, Alessandro Martini, Silvano Prosser, Nadia Giarbini, Christian Streitberger.   

Abstract

OBJECTIVE: To assess the functional results of the Vibrant Soundbridge (VBS) placed on the round window (RW) in patients with mixed hearing loss. STUDY
DESIGN: Retrospective evaluation of functional hearing, with measurements performed 7 to 9 months postoperatively. SETTINGS: Two tertiary referral hospital centers.
SUBJECTS: Twelve individuals with mixed severe hearing loss associated with chronic suppurative otitis media and otosclerosis. INTERVENTION: Surgical placement of the VBS mechanical effector in close contact with the RW membrane to directly drive the inner ear fluids. MAIN OUTCOME MEASUREMENT: Functional hearing gain as analyzed via pure-tone audiometry and speech audiometry with VBS off and on in quiet and in noise.
RESULTS: We observed a mean gain of 37.5 dB (0.5-4 kHz) with wide individual differences. The overall gain is mainly due to the air-bone gap recovery, whereas a further 12-dB mean improvement of air-conducted threshold is evident at 2 kHz. The speech reception threshold in quiet shows a mean gain of 24 dB, whereas in noise, it requires a signal-to-noise ratio 7 to 13 dB greater than normal-hearing controls. All patients are daily users of their VBS device.
CONCLUSION: A middle ear implant capable of directly driving the cochlear fluids seems to be a promising alternative for individuals with a severe to profound mixed hearing loss. However, variability in hearing recovery is great, likely reflecting variability in responsiveness of the cochlea associated with chronic pathologic findings and, possibly, variability in how the VBS effector interfaces with the RW. Modifying the shape of the VBS effector can improve the mechanical coupling to the RW to better exploit the device's amplification power.

Entities:  

Mesh:

Year:  2009        PMID: 19180678     DOI: 10.1097/mao.0b013e318180a495

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  35 in total

1.  Results and complications of the Baha system (bone-anchored hearing aid).

Authors:  G Ricci; A Della Volpe; M Faralli; F Longari; M Gullà; N Mansi; A Frenguelli
Journal:  Eur Arch Otorhinolaryngol       Date:  2010-06-10       Impact factor: 2.503

2.  Round window membrane implantation with an active middle ear implant: a study of the effects on the performance of round window exposure and transducer tip diameter in human cadaveric temporal bones.

Authors:  Stéphane Tringali; Kanthaiah Koka; Arnaud Deveze; N Julian Holland; Herman A Jenkins; Daniel J Tollin
Journal:  Audiol Neurootol       Date:  2010-02-11       Impact factor: 1.854

3.  Complications after round window vibroplasty.

Authors:  Maurizio Barbara; Luigi Volpini; Edoardo Covelli; Martina Romeo; Chiara Filippi; Simonetta Monini
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-03-27       Impact factor: 2.503

4.  Vibrant SoundBridge application to middle ear windows versus conventional hearing aids: a comparative study based on international outcome inventory for hearing aids.

Authors:  Ahmet Atas; Hakan Tutar; Bulent Gunduz; Yıldırım A Bayazıt
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-02-12       Impact factor: 2.503

5.  Oval and round window vibroplasty: a comparison of hearing results, risks and failures.

Authors:  Andrea Canale; Federico Dagna; Claudia Cassandro; Pamela Giordano; Federico Caranzano; Michelangelo Lacilla; Roberto Albera
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-10-11       Impact factor: 2.503

6.  Hearing rehabilitation with single-stage bilateral vibroplasty in a child with Franceschetti syndrome.

Authors:  Sona Sargsyan; Torsten Rahne; Sabrina Kösling; Gerburg Eichler; Stefan K Plontke
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-02-07       Impact factor: 2.503

7.  Measurement of basilar membrane motion during round window stimulation in guinea pigs.

Authors:  Yongzheng Chen; Xiying Guan; Tianyu Zhang; Rong Z Gan
Journal:  J Assoc Res Otolaryngol       Date:  2014-08-01

Review 8.  [The Vibrant Soundbridge as an active implant in middle ear surgery].

Authors:  T Beleites; M Bornitz; M Neudert; T Zahnert
Journal:  HNO       Date:  2014-07       Impact factor: 1.284

9.  The positional relationship between facial nerve and round window niche in patients with congenital aural atresia and stenosis.

Authors:  Keguang Chen; Huiying Lyu; Youzhou Xie; Lin Yang; Tianyu Zhang; Peidong Dai
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-03-12       Impact factor: 2.503

10.  Performance considerations of prosthetic actuators for round-window stimulation.

Authors:  Hideko Heidi Nakajima; Saumil N Merchant; John J Rosowski
Journal:  Hear Res       Date:  2009-11-23       Impact factor: 3.208

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.