Literature DB >> 19180657

Catheter-based treatment with the Amplatzer devices in alien positions.

Gunther Fischer1, Bjarne Smevik, Juan Carlo Ramiro Novoa, Francisco Javier Ozores Suáres, Hans Heiner Kramer, Per G Bjørnstad.   

Abstract

OBJECTIVE: Case reports of patients with Amplatzer devices used in positions they have not been designed for have repeatedly been published.
BACKGROUND: We pooled our results to review the array of such interventions.
METHODS: Retrospective study of twelve patients presented with eleven different problems: Two patients had hepatoatrial communication after Fontan's operation, two had different types of fistulae, two had surgical shunts, two had different anomalous systemic venous drainage, and two had an open arterial duct, one huge and hypertensive, the other with a very shallow ampulla. Finally, in one, a pulmonary artery was closed as a preparation for a bidirectional cavo-pulmonary anastomosis.
RESULTS: All communications were completely closed. As only complication a lesion of the anonymous vein causing haematothorax was noted when trying to access through a Gore-Tex membrane after puncture.
CONCLUSIONS: Such applications of Amplatzer devices in alien positions are estimated to be needed in 1 in 5-700 catheterizations. The results are good and the complications few. Copyright 2009 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2009        PMID: 19180657     DOI: 10.1002/ccd.21899

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  1 in total

1.  Early postoperative interventional ASD-closure for severe atrial right to left shunt in a neonate with common arterial trunk.

Authors:  Daniel Dilber; Andreas Eicken; John Hess
Journal:  Croat Med J       Date:  2013-08       Impact factor: 1.351

  1 in total