Quantitative analysis of pulsatile flow contribution to ultrafiltration.

Ki Moo Lim1, Joong Yull Park, Jung Chan Lee, Jeong Chul Kim, Byoung Goo Min, Eung Taeck Kang, Eun Bo Shim.   

Abstract

We evaluated the quantitative contribution of pulsatile flow to ultrafiltration (UF) in terms of fluid power, membrane stretch, and reduction of membrane layering. An in vitro comparison of the UF rate using pulsatile and roller pumps was performed with distilled water and bovine whole blood. The mean transmembrane pressure (TMPm) and UF rate were higher with the pulsatile pump for the same mean flow rate: 6.6 mm Hg and 21.1 mL/min higher on average for distilled water and 34.2 mm Hg and 31.4 mL/min higher on average for blood. The average UF rate was 8.4 mL/min higher with the pulsatile pump for the same TMPm with bovine blood. However, the relationship between the UF rate and the TMPm was independent of the flow configuration for distilled water. We showed that the higher UF rate in the pulsatile pump is mainly due to greater fluid power and reduction of membrane layering, while the membrane stretch was not an important factor.

Entities:  

Mesh:

Year:  2009        PMID: 19178443     DOI: 10.1111/j.1525-1594.2008.00683.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094