Orthogonal pin construct versus parallel uniplanar pin constructs for pelvic external fixation: a biomechanical assessment of stiffness and strength.

Michael T Archdeacon1, Sameh Arebi, T Toan Le, Rene Wirth, Roland Kebel, Mayur Thakore.   

Abstract

OBJECTIVE: The purpose of this study is to compare the structural stiffness of an orthogonal pelvic external fixator pin construct with 2 different parallel external fixator pin constructs in a simulated bone model. HYPOTHESIS: An orthogonal pelvic external fixator pin construct would be significantly more stiff than a parallel pin construct when loaded in-plane under similar conditions.
DESIGN: Thirty synthetic pelvic bone models were configured with orthogonal pins (group 1), parallel iliac crest pins (group 2), or parallel supra-acetabular pins (group 3). Specimens were loaded either in-plane (flexion/extension moment) or out-of-plane (internal/external rotation moment) to assess construct stiffness.
SETTING: Orthopaedic industry mechanical testing laboratory (Stryker Orthopedics, Mahwah, NJ). INTERVENTION: Single load cycle to failure with load application modified to assess stiffness both in-plane and out-of-plane with the pin constructs. MAIN OUTCOME MEASURE: Pelvic external fixation pin construct stiffness.
RESULTS: Stiffness for in-plane loading was 150.2 +/- 51.2 N/mm for the orthogonal pin construct, 105.0 +/- 46.9 N/mm for the iliac crest pin construct, and 104.7 +/- 20.7 N/mm for the supra-acetabular pin construct. Pairwise comparisons demonstrated that the difference was significant (P < 0.05) between groups 1 and 2 and groups 1 and 3 but not between groups 2 and 3. Stiffness for out-of-plane loading was 49.6 +/- 3.4 N/mm for the orthogonal pin construct, 53.9 +/- 3.5 N/mm for the iliac crest pin construct, and 100.6 +/- 4.3 N/mm for the supra-acetabular pin construct, with significant differences (P < 0.05) between groups 1 and 3 and groups 2 and 3 but not between groups 1 and 2.
CONCLUSION: An orthogonal pelvic external fixator pin construct produced a significantly stiffer construct for in-plane loading (flexion/extension moment) compared with either parallel pin construct; however, a parallel supra-acetabular pin construct was stiffer for out-of-plane loading.

Mesh:

Year:  2009        PMID: 19169101     DOI: 10.1097/BOT.0b013e3181962e0e

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  4 in total

1.  Achievable pin spanning angulation in anterosuperior pelvic external fixation.

Authors:  Benjamin Young; Drayton Daily; Clarence Kee; Kevin Perry; Massimo Max Morandi; R Shane Barton; Giovanni F Solitro
Journal:  Eur J Orthop Surg Traumatol       Date:  2020-05-01

2.  [Application of individualized transiliac crest nail-grafting guide plate in deep pelvic external fixator implantation].

Authors:  Cheng Tang; Bin Liang; Qingqiang Yao; Shuai Liu; Yan Xu; Liming Wang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2019-09-15

3.  A novel fixation system for acetabular quadrilateral plate fracture: a comparative biomechanical study.

Authors:  Guo-Chun Zha; Jun-Ying Sun; Sheng-Jie Dong; Wen Zhang; Zong-Ping Luo
Journal:  Biomed Res Int       Date:  2015-01-31       Impact factor: 3.411

4.  A feasibility study of individual 3D-printed navigation template for the deep external fixator pin position on the iliac crest.

Authors:  Bin Liang; Qiang Chen; Shuai Liu; Shuo Chen; Qingqiang Yao; Bo Wei; Yan Xu; Cheng Tang; Liming Wang
Journal:  BMC Musculoskelet Disord       Date:  2020-07-21       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.