Literature DB >> 19162873

A study of composite electrode-tissue impedance.

R L Robinson1, J L Davidson, P Wright, C J D Pomfrett, H McCann.   

Abstract

Electrical impedance tomography has been used in the medical environment for many years. Recently it has been used for brain function imaging. This type of technology uses a large number of electrodes on the scalp, examples in the literature ranging from 8 to 64. In this paper, the composite impedance of the electrode in contact with tissue is investigated. An experimental method has been developed to enable the use of a commercial impedance analyzer in compliance with medical safety standards. The magnitude and long-term stability of the composite impedance of the electrode-tissue system is found to be greatly dependent on the type of electrodes used.

Entities:  

Mesh:

Year:  2008        PMID: 19162873     DOI: 10.1109/IEMBS.2008.4649370

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  2 in total

1.  Scalp-Mounted Electrical Impedance Tomography of Cerebral Hemodynamics.

Authors:  Taweechai Ouypornkochagorn; Nataša Terzija; Paul Wright; John L Davidson; Nick Polydorides; Hugh McCann
Journal:  IEEE Sens J       Date:  2022-01-21       Impact factor: 4.325

Review 2.  Advances in electrical impedance tomography-based brain imaging.

Authors:  Xi-Yang Ke; Wei Hou; Qi Huang; Xue Hou; Xue-Ying Bao; Wei-Xuan Kong; Cheng-Xiang Li; Yu-Qi Qiu; Si-Yi Hu; Li-Hua Dong
Journal:  Mil Med Res       Date:  2022-02-28
  2 in total