BACKGROUND: Defining the stem insertion end point during total hip replacement still relies on the surgeon's feeling. When a custom-made stem prosthesis with an optimal fit into the femoral canal is used, the risk of per-operative fractures is even greater than with standard prostheses. Vibration analysis is used in other clinical settings and has been tested as a means to detect optimal stem insertion in the laboratory. The first per-operative use of vibration analysis during non-cemented custom-made stem insertion in 30 patients is reported here. MATERIALS AND METHODS: Thirty patients eligible for total hip replacement with uncemented stem prosthesis were included. The neck of the stem was connected with a shaker that emitted white noise as excitation signal and an impedance head that measured the frequency response. The response signal was sent to a computer that analyzed the frequency response function after each insertion phase. A technician present in the operating theatre but outside the laminated airflow provided feed-back to the surgeon. RESULTS: The correlation index between the frequency response function measured during the last two insertion hammering sessions was >0.99 in 86.7% of the cases. In four cases the surgeon stopped the insertion procedure because of a perceived risk of fracture. Two special cases illustrating the potential benefit of per-operative vibration analysis are described. CONCLUSIONS: The results of intra-operative vibration analysis indicate that this technique may be a useful tool assisting the orthopaedic surgeon in defining the insertion endpoint of the stem. The development of a more user-friendly device is therefore warranted.
BACKGROUND: Defining the stem insertion end point during total hip replacement still relies on the surgeon's feeling. When a custom-made stem prosthesis with an optimal fit into the femoral canal is used, the risk of per-operative fractures is even greater than with standard prostheses. Vibration analysis is used in other clinical settings and has been tested as a means to detect optimal stem insertion in the laboratory. The first per-operative use of vibration analysis during non-cemented custom-made stem insertion in 30 patients is reported here. MATERIALS AND METHODS: Thirty patients eligible for total hip replacement with uncemented stem prosthesis were included. The neck of the stem was connected with a shaker that emitted white noise as excitation signal and an impedance head that measured the frequency response. The response signal was sent to a computer that analyzed the frequency response function after each insertion phase. A technician present in the operating theatre but outside the laminated airflow provided feed-back to the surgeon. RESULTS: The correlation index between the frequency response function measured during the last two insertion hammering sessions was >0.99 in 86.7% of the cases. In four cases the surgeon stopped the insertion procedure because of a perceived risk of fracture. Two special cases illustrating the potential benefit of per-operative vibration analysis are described. CONCLUSIONS: The results of intra-operative vibration analysis indicate that this technique may be a useful tool assisting the orthopaedic surgeon in defining the insertion endpoint of the stem. The development of a more user-friendly device is therefore warranted.
Authors: Jonathan C J Wei; Willem H A Crezee; Hilda Jongeneel; Tobias S A De Haas; Wesley L A Kool; Bryan J Blaauw; Jenny Dankelman; Tim Horeman Journal: Sensors (Basel) Date: 2022-02-18 Impact factor: 3.576