Ventilator autocycling due to an endotracheal tube cuff leak.

R J Schwab1, J S Schnader.   

Abstract

Ventilator autocycling can occur with any ventilator if the sensitivity is improperly set or if a gas leak exists in the respiratory system which creates a negative change in proximal airway pressure. We report a case of ventilator autocycling in a paralyzed patient secondary to an endotracheal cuff leak which was misconstrued as assisted ventilation. We believe this is the first report of autocycling due to a cuff leak.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1914588     DOI: 10.1378/chest.100.4.1172

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  4 in total

1.  Ventilator autocycling and delayed recognition of brain death.

Authors:  William T McGee; Patrick Mailloux
Journal:  Neurocrit Care       Date:  2011-04       Impact factor: 3.210

2.  Cardiac Pacing a Rare Cause of Ventilator Auto Triggering.

Authors:  M Nithish Mukunthan; Vimal Bhardwaj
Journal:  Indian J Crit Care Med       Date:  2022-05

3.  Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling - a case series and literature review.

Authors:  Allen B Shoham; Bhavesh Patel; Francisco A Arabia; Michael J Murray
Journal:  J Cardiothorac Surg       Date:  2010-05-17       Impact factor: 1.637

4.  Home monitoring of daytime mouthpiece ventilation effectiveness in patients with neuromuscular disease.

Authors:  Julie Nardi; Karl Leroux; David Orlikowski; Hélène Prigent; Frédéric Lofaso
Journal:  Chron Respir Dis       Date:  2015-12-23       Impact factor: 2.444