Malte T Ahlers1, Thomas Bolz, Achim Bahr, Josef Ammermüller.
Abstract
Two exposure systems were developed for the measurement of retinal ganglion cell responses to light under the influence of pulsed high-frequency electromagnetic fields. Exposure characteristics were determined numerically for the GSM standards (900/1,800 MHz) and the UMTS standard (1,966 MHz) with specific absorption rates, averaged over the region of interest, of 0.02, 0.2, 2 und 20 W kg(-1). Extracellular multi- and single unit recordings of light responses from several retinal ganglion cells per retina could be obtained in these exposure systems on a regular basis, using two recording electrodes simultaneously. With appropriate temperature control adjustment, maximal temperature deviations at exposure onset and offset were well below the range of +/-0.1 degrees C for all SAR values.Mesh:
Year: 2009 PMID: 19142653 DOI: 10.1007/s00411-008-0207-7
Source DB: PubMed Journal: Radiat Environ Biophys ISSN: 0301-634X Impact factor: 1.925