Literature DB >> 19141995

Dislocation of a double-lumen tunneled hemodialysis catheter causing dysfunction and bleeding.

B Haid1, E Zitt, E Meusburger, E Wöss, K Lhotta.   

Abstract

Dislocation of tunneled hemodialysis catheters is rare. This report describes displacement of a split catheter with the tip of the arterial limb outside and the venous limb inside the jugular vein. Use of the catheter with the limbs in inverse fashion caused severe hemorrhage along the catheter tunnel. To avoid this complication, a short subcutaneous catheter tunnel is recommended in adipose patients.

Entities:  

Mesh:

Year:  2009        PMID: 19141995     DOI: 10.1159/000190783

Source DB:  PubMed          Journal:  Blood Purif        ISSN: 0253-5068            Impact factor:   2.614


  1 in total

1.  Delayed malposition of a double-lumen hemodialysis catheter that caused hemorrhage and hypovolemic shock: A case report.

Authors:  I-Chen Chen; Shih-Chia Yang; Kuan-Ting Liu; Yen-Hung Wu
Journal:  Medicine (Baltimore)       Date:  2019-01       Impact factor: 1.817

  1 in total