Lead-induced tricuspid stenosis--successful management by balloon angioplasty.

Tarique Hussain1, William Brodie Knight, Karen A McLeod.   

Abstract

We describe a case of tricuspid stenosis in a young woman 11 years after endocardial ventricular lead implantation. The cause of the stenosis appears to have been perforation of the septal leaflet by the lead at the time of insertion. We further describe successful management with percutaneous balloon valvuloplasty without the need for lead removal.

Entities:  

Mesh:

Year:  2009        PMID: 19140926     DOI: 10.1111/j.1540-8159.2009.02189.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Tricuspid stenosis: An emerging disease in cardiac implantable electronic devices era. Case report and literature review.

Authors:  Aysha Husain; Faris Tufail Raja; Ahmed Fatallah; Bahaa Fadel; Aly Alsanei; Fahad Tufail Raja; Bandar AlGhamdi
Journal:  J Cardiol Cases       Date:  2017-04-14

2.  Severe Tricuspid Stenosis Secondary to Permanent Pacemaker Lead.

Authors:  Asif Khan; Ahmad Mustafa; Joanne Ling; James Lafferty
Journal:  J Med Cases       Date:  2022-08-19

3.  Percutaneous tricuspid valvotomy for pacemaker lead-induced tricuspid stenosis.

Authors:  Devendra V Patil; Ashish A Nabar; Girish R Sabnis; Milind S Phadke; Charan P Lanjewar; Prafulla G Kerkar
Journal:  Indian Heart J       Date:  2016-01-14