Literature DB >> 19123143

Mild topographic abnormalities that become more suspicious on Scheimpflug imaging.

A Wolf1, W Abdallat, A Kollias, S J Frohlich, M Grueterich, C A Lackerbauer.   

Abstract

PURPOSE: Although several screening methods exist, postoperative corneal ectasia after refractive surgery is a severe complication. One possibility for this might be the fact that screening methods may fail in detection of preoperative risk factors such as forme fruste keratoconus (FFKC).
METHODS: Retrospective evaluation of four cases that showed only mild changes of FFKC on placido-based topography but revealed indicative findings on Scheimpflug imaging (Pentacam).
RESULTS: While in placido-based topography evaluation of corneal topography did not show a clear FFKC, the evaluation of corneal topography on Scheimpflug imaging together with the data of spatial corneal thickness revealed distinctive FFKC in all cases presented.
CONCLUSIONS: Although both methods bear the risk of not detecting pre-existing FFKC, Scheimpflug imaging seems superior to placido-based corneal topography alone.

Entities:  

Mesh:

Year:  2009        PMID: 19123143     DOI: 10.1177/112067210901900102

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  1 in total

Review 1.  Biomechanics of corneal ectasia and biomechanical treatments.

Authors:  Cynthia J Roberts; William J Dupps
Journal:  J Cataract Refract Surg       Date:  2014-04-26       Impact factor: 3.351