Literature DB >> 19110796

Quantitative multi-modality imaging analysis of a bioabsorbable poly-L-lactic acid stent design in the acute phase: a comparison between 2- and 3D-QCA, QCU and QMSCT-CA.

Nico Bruining1, Shuzou Tanimoto, Masato Otsuka, Annick Weustink, Jurgen Ligthart, Sebastiaan de Winter, Carlos van Mieghem, Koen Nieman, Pim J de Feyter, Ron T van Domburg, Patrick W Serruys.   

Abstract

AIMS: To investigate if three-dimensional (3D) based quantitative techniques are comparable to each other and to explore possible differences with respect to the reference method of 2D-QCA in the acute phase and to study whether non-invasive MSCT could potentially be applied to quantify luminal dimensions of a stented coronary segment with a novel bioabsorable drug-eluting stent made of poly-l-lactic-acid (PLLA). METHODS AND
RESULTS: Quantitative imaging data derived from 16 patients enrolled at our institution in a first-in-man trial (ABSORB) receiving a biodegradable stent and who were imaged with standard coronary angiography and intravascular ultrasound were compared. Shortly, after stenting the patients also underwent a MSCT procedure. Standard 2D-QCA showed significant smaller stent lengths (p < 0.01). Although, the absolute measured stent diameters and areas by 2D-QCA tend to be smaller, the differences failed to be statistically different when compared to the 3D based quantitative modalities. Measurements made by non-invasive QMSCT-CA of implanted PLLA stents appeared to be comparable to the other 3D modalities without significant differences.
CONCLUSIONS: Three-dimensional based quantitative analyses showed similar results quantifying luminal dimensions as compared to 2D-QCA during an evaluation of a new bioabsorbable coronary stent design in the acute phase. Furthermore, in biodegradable stents made of PLLA, non-invasive QMSCT-CA can be used to quantify luminal dimensions.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19110796     DOI: 10.4244/eijv4i2a49

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  4 in total

1.  Quantitative multi-modality imaging analysis of a fully bioresorbable stent: a head-to-head comparison between QCA, IVUS and OCT.

Authors:  Juan Luis Gutiérrez-Chico; Patrick W Serruys; Chrysafios Girasis; Scot Garg; Yoshinobu Onuma; Salvatore Brugaletta; Héctor García-García; Gerrit-Anne van Es; Evelyn Regar
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-26       Impact factor: 2.357

2.  In vivo comparison of arterial lumen dimensions assessed by co-registered three-dimensional (3D) quantitative coronary angiography, intravascular ultrasound and optical coherence tomography.

Authors:  Shengxian Tu; Liang Xu; Jurgen Ligthart; Bo Xu; Karen Witberg; Zhongwei Sun; Gerhard Koning; Johan H C Reiber; Evelyn Regar
Journal:  Int J Cardiovasc Imaging       Date:  2012-01-20       Impact factor: 2.357

3.  Fusion of 3D QCA and IVUS/OCT.

Authors:  Shengxian Tu; Niels R Holm; Gerhard Koning; Zheng Huang; Johan H C Reiber
Journal:  Int J Cardiovasc Imaging       Date:  2011-01-25       Impact factor: 2.357

4.  Assessment of obstruction length and optimal viewing angle from biplane X-ray angiograms.

Authors:  Shengxian Tu; Gerhard Koning; Wouter Jukema; Johan H C Reiber
Journal:  Int J Cardiovasc Imaging       Date:  2009-09-18       Impact factor: 2.357

  4 in total