Literature DB >> 19101284

CardioWest (Jarvik) total artificial heart: a single-center experience with 42 patients.

Jean Christian Roussel1, Thomas Sénage, Olivier Baron, Christian Périgaud, Oussama Habash, Jean Christophe Rigal, Michèle Treilhaud, Jean Noel Trochu, Philippe Despins, Daniel Duveau.   

Abstract

BACKGROUND: When implanted in patients with biventricular failure, the CardioWest total artificial heart has asserted itself over time as a reliable bridge-to-transplant device that as yet is used by only a few international teams. The aim of this single-center retrospective study is to assess both the comorbidity and survival of patients awaiting heart transplants while receiving circulatory support with a CardioWest total artificial heart.
METHODS: From 1990 to December 2006, 42 patients received a CardioWest total artificial heart at our center. Mean age at the time of implantation was 45.7 +/- 9.5 years, and 40 patients (95%) were men. Idiopathic or dilated cardiomyopathy was diagnosed in 45.2% (n = 19) of the patients and ischemic cardiomyopathy in 42.8% (n = 18). Average body surface area was 1.9 +/- 0.22 m(2).
RESULTS: Duration of support was 1 to 292 days (mean, 101 +/- 86 days). Twelve patients died (28.5%) while receiving device support, and 30 patients (71.5%) underwent transplantation. Actuarial survival rates for the transplanted patients were 90% (n = 25), 81% (n = 14), and 76% (n = 10) at 1, 5, and 10 years, respectively. Causes of death during device support included multiorgan failure in 6 (50%), sepsis in 2, acute respiratory distress syndrome in 2, alveolar hemorrhage in 1, and other cause in 1. There were no device malfunctions that led to patient death. Adverse events included stroke in 3 patients (7%) and infections in 35 patients (85%) during support.
CONCLUSIONS: The CardioWest total artificial heart is an excellent bridge-to-transplant device for patients with biventricular failure. Our study demonstrates excellent safety, reliability, and efficiency. Exceptional outcome after transplantation underlines its capacity to aid in end-organ recovery.

Entities:  

Mesh:

Year:  2009        PMID: 19101284     DOI: 10.1016/j.athoracsur.2008.09.048

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  11 in total

Review 1.  [Total artificial heart].

Authors:  H Antretter; J Dumfarth; D Höfer
Journal:  Med Klin Intensivmed Notfmed       Date:  2015-08-22       Impact factor: 0.840

2.  Cardiovascular implantable device infections.

Authors:  George M Viola; Rabih O Darouiche
Journal:  Curr Infect Dis Rep       Date:  2011-08       Impact factor: 3.725

Review 3.  Mechanical circulatory support: devices, outcomes and complications.

Authors:  Carmelo A Milano; Alan A Simeone
Journal:  Heart Fail Rev       Date:  2013-01       Impact factor: 4.214

Review 4.  Cardiogenic shock in ACS. Part 2: Role of mechanical circulatory support.

Authors:  Stephen Westaby; Kyriakos Anastasiadis; George M Wieselthaler
Journal:  Nat Rev Cardiol       Date:  2012-01-10       Impact factor: 32.419

Review 5.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

6.  Progress on the design and development of the continuous-flow total artificial heart.

Authors:  Mariko Kobayashi; David J Horvath; Nicole Mielke; Akira Shiose; Barry Kuban; Mark Goodin; Kiyotaka Fukamachi; Leonard A R Golding
Journal:  Artif Organs       Date:  2012-07-02       Impact factor: 3.094

7.  Implantation of total artificial heart in congenital heart disease.

Authors:  Iki Adachi; David S L Morales
Journal:  J Vis Exp       Date:  2014-07-18       Impact factor: 1.355

8.  Experience with the SynCardia total artificial heart in a Canadian centre.

Authors:  Anthony Nguyen; Michel Pellerin; Louis P Perrault; Michel White; Anique Ducharme; Normand Racine; Michel Carrier
Journal:  Can J Surg       Date:  2017-12       Impact factor: 2.089

9.  Repair of left ventricular driveline tear in a SynCardia-total artificial heart patient.

Authors:  Sotirios Spiliopoulos; Magda Tenderich; Dilek Guersoy; Guenes Dogan; Reiner Koerfer; Gero Tenderich
Journal:  J Cardiothorac Surg       Date:  2014-01-07       Impact factor: 1.637

10.  Shear stress, energy losses, and costs: a resolved dilemma of pulsatile cardiac assist devices.

Authors:  Sayed Nour; Jia Liu; Gang Dai; Daniel Carbognani; Daya Yang; Guifu Wu; Qinmei Wang; Juan Carlos Chachques
Journal:  Biomed Res Int       Date:  2014-01-08       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.