The clinical reliability of automated cerebrospinal fluid cell counts on the Beckman-Coulter LH750 and Iris iQ200.

Lewis Glasser1, Carole A Murphy, Jason T Machan.   

Abstract

Although automation has improved the accuracy and precision of blood cell counts and is more rapid and less labor-intensive, cerebrospinal fluid (CSF) samples are still counted manually. We compared the IRIS iQ200 Body Fluids Module (Iris Diagnostics, Chatsworth, CA) and the Beckman-Coulter LH750 (Beckman-Coulter, Brea, CA) with manual counts and evaluated the impact of automation on the laboratory if clinically acceptable performance was to be maintained. Automated counts were compared with manual counts on 313 specimens. Clinical reliability was estimated using the weighted kappa coefficient and the impact of errors discussed in the context of a historic census of 3,653 samples spanning 19 months. Nucleated cell counts had a reliability of 0.73 for the LH750 and 0.84 for the iQ200. However, our results showed unacceptable rates of error at counts less than 200/microL (200 x 106/L) for the LH750 and less than 50/microL (50 x 106/L) for the iQ200, representing 94% and 83% of the census specimens, respectively. If clinical reliability is to be maintained, neither the LH750 nor iQ200 would have a significant impact on improving the efficiency of the laboratory because of the high percentage of low CSF cell counts.

Mesh:

Year:  2009        PMID: 19095566     DOI: 10.1309/AJCPCHHTZ6UDQ6WF

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  4 in total

1.  Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i.

Authors:  Sabrina Buoro; Sara Apassiti Esposito; MariaGrazia Alessio; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Improvement of cell counting method for Neubauer counting chamber.

Authors:  Minghao Zhang; Lingui Gu; Peihua Zheng; Zhixin Chen; Xinqi Dou; Qizhong Qin; Xiaozhong Cai
Journal:  J Clin Lab Anal       Date:  2019-08-30       Impact factor: 2.352

3.  Reducing sources of variance in experimental procedures in in vitro research.

Authors:  Igor Fischer; Maria Victoria Martinez-Dominguez; Daniel Hänggi; Ulf Kahlert
Journal:  F1000Res       Date:  2021-10-12

4.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.