[Calculation of the left ventricular ejection fraction. Comparison between 4 different instruments].

R Puchal Añé1, S Guirao Marín, A Domènech Vilardell, A Rodríguez Gassén, M T Bajén Lázaro, Y Ricart Brulles, M Murillo Sosa, J Martín-Comín.   

Abstract

AIM: The aim of this study was to analyse the relationship between left ventricular ejection fractions (EF) obtained using four different instruments.
MATERIAL AND METHODS: Eighty-five EF values were analysed. Gated acquisition was performed with the same gammacamera (SOPHA) using 99mTc-labelled red cells. Each EF was obtained using four processing systems: NXT (SOPHA), Entegra (Gems), Odyssey (Philips) and Esoft (Siemens), always working in automated mode. The paired student's t-test, Spearman correlation and Bland-Altman analysis were used to compare methods, and Deming regression was applied.
RESULTS: Mean values and standard deviations for each program were: NXT: 61 +/- 9; Entegra: 60 +/- 10; Odyssey: 60 +/- 9; Esoft: 60 +/- 10. Although no significant differences were found as a whole and the values were linearly related, the methods are not interchangeable.
CONCLUSIONS: The same program should be used in the follow-up of each patient, which is now easily achievable by means of the DICOM standard.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19094900

Source DB:  PubMed          Journal:  Rev Esp Med Nucl        ISSN: 0212-6982


  1 in total

1.  Comparison of estimates of left ventricular ejection fraction obtained from gated blood pool imaging, different software packages and cameras.

Authors:  Rachelle Steyn; John Boniaszczuk; Theodore Geldenhuys
Journal:  Cardiovasc J Afr       Date:  2014 Mar-Apr       Impact factor: 1.167

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.