Literature DB >> 19094168

Repeated measures of macular pigment optical density to test reproducibility of heterochromatic flicker photometry.

Stefan Hagen1, Ilse Krebs, Carl Glittenberg, Susanne Binder.   

Abstract

PURPOSE: To report the reproducibility of macular pigment optical density (MPOD) values assessed with heterochromatic flicker photometry (HFP) in healthy individuals.
METHODS: Twenty-four volunteers from our department underwent MPOD testing of both eyes by flicker photometry on three separate occasions. To test reproducibility of MPOD, the coefficient of variance was calculated separately for right and left eyes. In addition, we investigated MPOD averages of right and left eyes and interocular correlations (Pearson's r) at every visit.
RESULTS: The mean MPODs at the first visit were 0.61 +/- 0.24 and 0.72 +/- 0.27 in right and left eyes, respectively. Mean values of 0.58 +/- 0.29 and 0.60 +/- 0.21 (second visit) and 0.62 +/- 0.27 and 0.63 +/- 0.24 (third visit) were assessed for right and left eyes, respectively. Differences of the mean values between eyes were not significant. Correlations were weak at visits one and two (r = 0.49, p < 0.014 and r = 0.43, p < 0.038, respectively) and moderate at visit three (r = 0.58, p < 0.003). The coefficients of variance were 36.1% and 23% for right and left eyes, respectively.
CONCLUSION: Our mean MPODs are higher and the interocular correlations weaker compared to the literature. The coefficient of variance in both eyes is high and does not imply good reproducibility of obtained MPOD values.

Mesh:

Substances:

Year:  2008        PMID: 19094168     DOI: 10.1111/j.1755-3768.2008.01418.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  4 in total

1.  Macular pigment optical density measurements: evaluation of a device using heterochromatic flicker photometry.

Authors:  R de Kinkelder; R L P van der Veen; F D Verbaak; D J Faber; T G van Leeuwen; T T J M Berendschot
Journal:  Eye (Lond)       Date:  2010-11-05       Impact factor: 3.775

2.  The use of heterochromatic flicker photometry to determine macular pigment optical density in a healthy Australian population.

Authors:  Robin G Abell; Alex W Hewitt; Marko Andric; Penelope L Allen; Nitin Verma
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-01-05       Impact factor: 3.117

Review 3.  Measuring macular pigment optical density in vivo: a review of techniques.

Authors:  Olivia Howells; Frank Eperjesi; Hannah Bartlett
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-01-08       Impact factor: 3.117

4.  Short- and mid-term repeatability of macular pigment optical density measurements using spectral fundus reflectance.

Authors:  Nikolaus Dragostinoff; René Marcel Werkmeister; Semira Kaya; Günther Weigert; Berthold Pemp; Stefan Sacu; Gerhard Garhöfer; Ursula Schmidt-Erfurth; Leopold Schmetterer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-02-25       Impact factor: 3.117

  4 in total