Literature DB >> 19092456

The influence of central corneal thickness and corneal curvature radius on the intraocular pressure as measured by different tonometers: noncontact and goldmann applanation tonometers.

Yukinori Harada1, Naofumi Hirose, Toshiaki Kubota, Akihiko Tawara.   

Abstract

PURPOSE: The aim of this study was to examine whether or not central corneal thickness (CCT) and corneal curvature radius (CCR) influence the concordance of intraocular pressure (IOP) readings taken with a noncontact tonometer (NCT) and a Goldmann applanation tonometer (GAT).
METHODS: Eighty-seven eyes from 87 patients without any histories of intraocular operation were enrolled in the present study. We calculated the difference in IOP readings between the 2 methods (NCT-GAT), and then examined the relationship between the IOP readings, and CCT and CCR using a linear regression line.
RESULTS: IOP measured by both NCT (r=0.515, P<0.0001) and GAT (r=0.237, P<0.05) was significantly correlated with CCT. NCT readings were significantly higher in the thicker group (CCT>or=526.9 microm) than in the thinner group (CCT<526.9 microm) (P<0.005). GAT readings had no difference between the thicker and thinner groups. IOP measured by GAT (r=-0.261, P<0.05), but not that measured by NCT (r=-0.111, P=0.3042), showed a significant negative correlation with CCR. An IOP value of (NCT-GAT) had a significant positive correlation with not only CCT (r=0.666, P<0.0001) but also CCR (r=0.227, P<0.05). The value of (NCT-GAT) was significantly higher in the thicker group than in the thinner group (P<0.005).
CONCLUSIONS: NCT can be more affected by CCT than GAT. CCR could influence the measurements of GAT, but not those by NCT. CCT and CCR can therefore influence the discordance of IOP readings taken with NCT and GAT.

Entities:  

Mesh:

Year:  2008        PMID: 19092456     DOI: 10.1097/IJG.0b013e3181634f0f

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  12 in total

1.  Purinergic dysregulation causes hypertensive glaucoma-like optic neuropathy.

Authors:  Youichi Shinozaki; Kenji Kashiwagi; Kazuhiko Namekata; Akiko Takeda; Nobuhiko Ohno; Bernard Robaye; Takayuki Harada; Takeshi Iwata; Schuichi Koizumi
Journal:  JCI Insight       Date:  2017-10-05

2.  Relationship of central corneal thickness to postural IOP changes in patients with and without glaucoma in southern India.

Authors:  Stephen J Moster; Ghasem Fakhraie; Rengaraj Venketesh; Mark L Moster; Yuanjun Zhao; Marlene R Moster
Journal:  Int Ophthalmol       Date:  2012-05-11       Impact factor: 2.031

3.  Two-position measurement of intraocular pressure by PT100 noncontact tonometry in comparison with Goldmann tonometry.

Authors:  Kelechi C Ogbuehi; John C Chijuka; Uchechukwu L Osuagwu
Journal:  Clin Ophthalmol       Date:  2011-09-06

4.  Corneal Thickness Profile and Associations in Chinese Children Aged 7 to 15 Years Old.

Authors:  Yingyan Ma; Xiaofeng Zhu; Xiangui He; Lina Lu; Jianfeng Zhu; Haidong Zou
Journal:  PLoS One       Date:  2016-01-11       Impact factor: 3.240

5.  Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy.

Authors:  Mahmoud Jabbarvand; Farshad Askarizadeh; Mohamad Reza Sedaghat; Hadi Ghadimi; Bahram Khosravi; Mohammad Aghazadeh Amiri; Foroozan Narooie-Noori
Journal:  J Ophthalmic Vis Res       Date:  2017 Oct-Dec

6.  Refinement of Intraocular Pressure Measurements made by Ocular Response Analyzer following Laser in Situ Keratomileusis using M2 90 Moria Microkeratome for Egyptian Myopic and Astigmatic Patients.

Authors:  Tamer Adel Refai; Olfat A Hassanin
Journal:  Electron Physician       Date:  2016-12-25

7.  Intraocular pressure in a cohort of healthy eastern European schoolchildren: variations in method and corneal thickness.

Authors:  Patrycja Krzyżanowska-Berkowska; Magdalena Asejczyk-Widlicka; Barbara Pierscionek
Journal:  BMC Ophthalmol       Date:  2012-12-02       Impact factor: 2.209

8.  The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes.

Authors:  Jitendra Jethani; Paaraj Dave; Monica Jethani; Yogesh Desai; Purvi Patel
Journal:  Saudi J Ophthalmol       Date:  2015-11-19

9.  Changes of intraocular pressure and refractive status in children following cycloplegic refraction with 1% cyclopentolate and 1% tropicamide.

Authors:  Kuo-Chi Hung; Hsiu-Mei Huang; Pei-Wen Lin
Journal:  Taiwan J Ophthalmol       Date:  2015-08-24

10.  The influence of the tonometer position on canine intraocular pressure measurements using the Tonovet® rebound tonometer.

Authors:  Juliana Kravetz de Oliveira; Fabiano Montiani-Ferreira; David L Williams
Journal:  Open Vet J       Date:  2018-03-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.