The compatibility of denture cleansers and resilient liners.

Luciana Valadares Oliveira1, Marcelo Ferraz Mesquita, Guilherme Elias Pessanha Henriques, Rafael Leonardo Xediek Consani, Wagner Sotero Fragoso.   

Abstract

PURPOSE: Difficulty in cleaning resilient denture liners remains a material disadvantage. The purpose of the present study was to evaluate the effect of denture cleansers on hardness of resilient liner materials.
MATERIALS AND METHODS: Three resilient liners, Luci Sof (Dentsply), Molloplast-B (Dentax), and Sofreliner (Tokuyama), and two denture cleansers, Efferdent (Warner-Lamber), and 0.5% alkaline hypochlorite preparation were used. Twenty specimens of each material were prepared, measuring 25X15X3mm. Two denture cleansing approaches were used: 1) alkaline hypochlorite, for 20 minutes; 2) alkaline peroxide, for 30 minutes. This procedure was repeated 8 times a day, during 90 days. The specimens were evaluated before and after 360 and 720 cycles, to simulate 1 and 2 years of clinical cleaning procedures, respectively. The Shore A hardness was evaluated in a durometer (Teclock GS-709A), with a penetrating load of 10N for 1 second. Any macroscopic changes, such as loss of color or alteration in surface texture were recorded by one observer. All numeric data were subject to ANOVA with repeated measures followed by Tukey's test (alpha= 0.05).
RESULTS: All materials were significantly different, independently to time and treatment. Initially, Luci Sof and Sofreliner immersed in either hypochlorite or peroxide increased the hardness mean values significantly. These hardness mean values decreased significantly after 720 cycles. Molloplast-B showed no significant difference after the treatments, in any time.
CONCLUSIONS: Denture cleansers had no effect on hardness of the resilient denture liners evaluated after 2 years of in vivo simulated conditions of hygiene. Sofreliner was the smoothest material before and after all treatments.

Entities:  

Year:  2006        PMID: 19089278      PMCID: PMC4327488          DOI: 10.1590/s1678-77572006000400014

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

Even though acrylic resin is commonly used for complete denture bases, patients often prefer resilient denture liners to conventional hard denture bases20. The use of resilient liners is designed to distribute functional and nonfunctional stresses more evenly and to have a dampening effect because of elastic behavior4. These properties make resilient liners useful for treating patients with atrophic ridge or resorption, bony undercuts, bruxism, congenital or acquired oral defects requiring obturation, xerostomia and dentures opposing natural teeth. Commercially available products include rubbery acrylic-, silicone-, fluoric- and olefin-type materials. However, there are some disadvantages to the use of these materials. It has been shown that resilient liners are easily colonized by Candida albicans 12, which may cause an oral pathologic condition known as denture stomatitis when associated with poor oral and denture hygiene1. Prosthetic cleansing that removes Candida albicans is a necessary and important factor in preventing non-traumatic causes of denture stomatitis18. Consequently, the greatest disadvantage of the resilient lining material is the difficulty to maintain it clean10. Routine methods commonly used for denture cleaning include the use of immersion cleansers and brushing2,7. Brushing with or without an abrasive cleanser is also very effective in removing plaque; consequently, this cleansing method is very common2. It is still questionable whether brushing causes abrasion of resilient lining materials10, 17. While Wright20 and Schmidt and Smith17 observed no evidence that resilient liners are more difficult to clean than conventional dentures, neither were there any evidence of abrasion or wear of its surface, Makila and Honka10 found wear of the material when submitted to brushing. The immersion type denture cleansers may be defined as alkaline peroxides and hypochlorites, acids, disinfectants, and enzymes2. Ideally these solutions should be effective in removing stains and deposits from the denture; it should be simple to use and compatible with all denture base materials7. Although chemical denture cleansers have been considered to be an efficacious method to prevent C. albicans colonization and denture plaque formation6, 13, it is cited that a daily use of denture cleansers can affect the properties of both denture acrylic resin and resilient liners8, 11. Thus, dentists should choose denture cleansers by taking into account the compatibility of denture cleansers with resilient liners on both materials and biological aspects. Resilient liners are made of materials from several chemical families. These materials undergo chemical changes over time as patients immerse their dentures in either the aqueous environment of the mouth or, when not in use, in tap water or denture cleansers. Since hardening of the material is one of the major reasons for failure of some resilient liners, and difficulty in cleaning these materials remains their disadvantage, the purpose of the present study was to investigate the effect of immersion denture cleansers on the hardness of three resilient liners. The hypothesis tested was that the immersion cleansers would not affect the hardness of the resilient liners evaluated.

MATERIAL AND METHODS

The resilient liner materials and the denture cleansers used in this study are listed in Table 1. The resilient liners, Luci Sof ®, Molloplast-B® and Sofreliner® were select because they are commonly used by clinicians. The test solutions consisted of one commercially available denture cleanser and one alkaline hypochlorite solution.
TABLE 1

Resilient denture liners and denture cleansers

BrandMaterialBatch N.Manufacturer
Luci SofHeat-curing silicone990726ADentsply International Inc., York, PA - USA
Molloplast-BHeat-curing silicone011262Dentax-Gmbh & Co. KG, Ettlingen - Germany.
SofrelinerChemical-curing SiliconeU46973Tokuyama Dental Corp. Tokyo - Japan.
EfferdentAlkaline PeroxideBH 0783VPfizer Consumer Health Care, Morris Plains, NJ - EUA
AlkalineHypochlorite0.5% Alkaline Hypochlorite011262Medicinallis - Farmácia de Manipulação e Homeopatia, Piracicaba, SP – Brazil.
Thirty specimens (24 X 14 X 3 mm) of each material were prepared by investing dies in a standard metal dental flask (Uraby; DLC, São Paulo, Brazil). The thickness of the tested specimens chosen for this study was, as far as practical, representative of the clinical situation16, 17. The dies were invested in flexible silicone rubber (Zetalabor; Zhermack, Badia Polesine, Rovigo, Italy) to allow easy removal of the processed specimens from the flask. The flask was then completed with Type III dental stone (Herodent Soli-Rock; Vigodent, Rio de Janeiro, Brazil). After the complete setting of the gypsum, the two halves of the flask were separated and the dies removed. A separated medium (Cel Lac; S. S. White, Rio de Janeiro, Brazil) was applied to the exposed areas of the mold. Specimens were made by processing the resilient denture liners in the resulting mold according to the manufactures' instructions. Luci Sof® was processed by heating the flask in a water bath at 100 °C for 2 ½hours. Molloplast-B® was processed in a water bath at 100 °C for 2 hours. Sofreliner® was processed at room temperature. After the polymerization, specimens were removed from the flask, and any flash was trimmed with a sharp knife. The specimens were then removed and stored in water at 37 ° C for 24 hours, and randomly divided into 2 groups. The immersion periods were chosen to resemble normal domestic usage. The alkaline peroxide solution was prepared by adding one tablet of Efferdent® to 200ml tap water (equivalent to the dilution recommended by the manufacturer) which was at initial temperature of 37 °C. The solution changed from blue to clear, and after 15 minutes, the specimens were washed thoroughly with tap water and distilled water. The specimens were immersed into 200 ml of 0.5% alkaline hypochlorite solution at initial temperature of 37 °C for 20 minutes. Fresh solutions were used for each immersion. Ten specimens of each material were immersed into the solution of each denture cleanser 8 times each day for a total of 720 treatments extending over a period of 90 days. Between the immersion periods, the specimens were stored in distilled water at 37 °C for the remainder of the 24 hours period.

Shore A Hardness Test

Hardness was measured using a Shore Durometer (GS-709, Teclock – Osaka, Japan), under a load of 10N for 1 second. The instrument consists of a blunt-pointed indenter attached to a scale by lever arrangement with a recording scale from 0 to 100 Shore A units. The more the indenter penetrates the specimen, the lower the hardness value is. The shaft was lowered onto the middle of the specimen, and the highest reading was noted. Five readings were obtained for each specimen, and the final hardness mean value calculated for each specimen. One representative specimen was selected of each group. The data from the each assay were subjected to Analysis of Variance with repeated measures and Tukey test (5%).

Visual Assessment

All specimens were examined visually on days 360 and 720 and were compared with the controls so that one observer could note any macroscopic changes such as loss of color or alteration in surface texture. Any changes were recorded on a grading system of slight, moderate or marked3. All numeric data were submitted to Analysis of Variance followed by Tukey's test (α= 0.05).

RESULTS

Table 2 lists the hardness mean values of the resilient liners subjected to the cleansing treatments. Before the immersion treatments, Sofreliner® presented the lowest hardness mean values, followed by Molloplast-B® and Luci Sof®. The denture cleansers treatment significantly increased the hardness mean values of Luci Sof® and Sofrelinef® after 360 immersion cycles. However, these values decreased significantly after 720 immersion cycles. The denture cleansers treatment did not significantly affect Molloplast-B® mean values, either after 360 or 720 immersion cycles.
TABLE 2

Mean hardness and ± standard deviation of denture materials before and after 360 and 720 cycles (n=10)

Denture CleanserMaterialControlTime 360720
PeroxideLS42.75 ± 2.02 Ab43.51 ± 1.11 Aa42.77 ± 2.05Ab
MO37.71 ± 3.16 Ba37.74 ± 2.60 Ba37.81 ± 3.37 Ba
SO28.76 ± 1.92 Cb31.50 ± 1.33 Ca28.76 ± 1.92Cb
HypochloriteLS43.62 ± 1.59 Ab44.46 ± 1.31Aa43.62 ± 1.59 Ab
MO37.34 ± 2.38 Ba37.53 2.73± Ba36.84 ± 2.98Ba
SO29.25 ± 1.99 Cb30.67 ± 1.44 Ca29.25 ± 1.99Cb

Means followed by the same upper case letter in a column (time and treatment) and the same lower case letter in a row (material and treatment) do not differ statistically by Tukey at a 5% probability level.

(p = 0.7440)

Means followed by the same upper case letter in a column (time and treatment) and the same lower case letter in a row (material and treatment) do not differ statistically by Tukey at a 5% probability level. (p = 0.7440) The changes observed after 360 and 720 are summarized in Table 3. Marked color loss was observed after 720 immersion cycles for all materials. The peroxide treatment promoted loss of surface sheen of all materials.
TABLE 3

Observed changes in resilient denture liners after 360 and 720 immersions in denture cleansers

TreatmentMaterial360720
HypochloriteLuci SofSlight colour lossSlight colour loss
Molloplast-BSlight colour loss; Slight loss of surface sheen.Marked colour loss; Marked loss of surface sheen.
SofrelinerSlight loss of surface sheen.Marked colour loss; Slight loss of surface sheen.
PeroxideLuci SofModerate colour lossMarked colour loss
Molloplast-BSlight colour loss; Moderate loss of surface sheen.Marked colour loss; Marked loss of surface sheen.
SofrelinerSlight loss of surface sheen.Moderate colour loss; Moderate loss of surface sheen.

DISCUSSION

Resilient liners undergo through changes over time as patients immerse their dentures in either aqueous environment of the mouth or, when not in use, in tap water or denture cleansers. Hardening of the material is one of the major reasons for failure of resilient liners9. Therefore, it is important to evaluate if denture cleansers can affect the hardness of the resilient liners. Initial measurements showed that Luci Sof® presented the highest hardness mean values, followed by Molloplast-B® and Sofrelinef®. Hardness mean values for the heat-cured materials (Luci Sof® and Molloplast-B®) were significantly greater than those for the self-curing one, in agreement with other studies9,14,15. This difference might be explained because of the polymerization mode of the materials as well as their composition. Increased processing temperatures were expected to result in a more complete polymerization reaction and thus a harder polymer15. Although Luci Sof® and Molloplast-B® are heat-cured materials, they have different polymerization periods. Luci Sof® was processed for 2^2 hours in boiling water, and Molloplast-B® was processed for 2 hours in boiling water, following the manufacturer's directions. This different period of processing resulted in a more complete polymerization reaction14, and this might lead to higher hardness mean values. Sofreliner® is a chemically cured material, and this processing type might result in incomplete initial polymerization and thus lower hardness values. Luci Sof® and Sofreliner® hardened significantly after 360 immersion cycles. This was likely due to continued polymerization of those, as cited by Dootz, et al.4 (1993). The materials softened during the remainder of the test period, approaching the initial hardness mean values by the end of the study period, in agreement with Kiat-Amnuay, et al.9 (2005), who tested resilient liners' hardness after water storage. Silicone-based materials do not contain plasticizer, but does contain filler16, and the absorption of water by the filler could lead to increased softness of the materials evaluated. The decrease of hardness mean values after immersion in denture cleansers occurred because of the absorption of water by the fillers. However, denture cleansers had no effect on Molloplast-B® mean hardness values, after either 360 or 720 cycles. Tan, et al.19 (2000) did not find significant differences in Molloplast-B® mean values after immersion treatment in either hypochlorite or peroxide. The authors suggest that this was resulted because of an arbitrary test period and a small experimental group. However, the present study evaluated a longer test period and twenty specimens, and the material presented a linear behavior during it. Although hypochlorite caused a marked loss of color in all materials tested, it is questionable whether this change would be of importance clinically. However, it is advisable to warn patients who are provided with a denture relined with these materials that although fading of the lining might occur, this does not indicate that the resilient liner is unsatisfactory. The resilient liners also showed some mirror surface change in the form of loss of sheen. Davenport, et al.3 (1986) suggests that this is the result of an increase in surface roughness. Plaque retention is increased in this situation. However, this is less likely to be of significance when the denture cleanser being used is effective in removing plaque from prostheses. Within the limitations of this study, Sofreliner® displayed the lowest hardness values before and after the immersion in denture cleansers. According to Gonzalez5 resilient liners should have a Shore A Durometer hardness of approximately 20 to 25 units and not change hardness with service. Only Sofreliner® presented values close to those suggested. However, it should be elucidated if this difference is clinically relevant. In addition, selection of a resilient liner cannot be based on any single property. Lower hardness values is a desirable property for these materials, but other factors, such as bond strength to denture base, lower surface roughness values, tear strength, water sorption and solubility, must also be considered when selecting these materials.

CONCLUSIONS

Under the conditions of this study, the following conclusions were drawn: Denture cleansers significantly increased Luci Sof® and Sofreliner® hardness mean values after 360 immersion cycles, and decreased significantly after 720 cycles. Denture cleansers had no effect on Molloplast-B® hardness mean values either after 360 or 720 immersion cycles. Sofreliner® was the smoothest material under the conditions of the present study, before and after the treatments.
  20 in total

1.  Changes in surface roughness and colour stability of soft denture lining materials caused by denture cleansers.

Authors:  C Jin; H Nikawa; S Makihira; T Hamada; M Furukawa; H Murata
Journal:  J Oral Rehabil       Date:  2003-02       Impact factor: 3.837

2.  The influence of water storage on durometer hardness of 5 soft denture liners over time.

Authors:  Sudarat Kiat-Amnuay; Lawrence Gettleman; Trakol Mekayarajjananonth; Zafrulla Khan; L Jane Goldsmith
Journal:  J Prosthodont       Date:  2005-03       Impact factor: 2.752

3.  Influence of curing method, sealer, and water storage on the hardness of a soft lining material over time.

Authors:  G L Polyzois; M J Frangou
Journal:  J Prosthodont       Date:  2001-03       Impact factor: 2.752

Review 4.  The significance of Candida albicans in denture stomatitis.

Authors:  E Budtz-Jörgensen
Journal:  Scand J Dent Res       Date:  1974

5.  A six-year retrospective study of Molloplast-B-lined dentures. Part II: Liner serviceability.

Authors:  W F Schmidt; D E Smith
Journal:  J Prosthet Dent       Date:  1983-10       Impact factor: 3.426

6.  The success and failure of denture soft-lining materials in clinical use.

Authors:  P S Wright
Journal:  J Dent       Date:  1984-12       Impact factor: 4.379

7.  Biofilm formation of Candida albicans on the surfaces of deteriorated soft denture lining materials caused by denture cleansers in vitro.

Authors:  H Nikawa; C Jin; S Makihira; H Egusa; T Hamada; H Kumagai
Journal:  J Oral Rehabil       Date:  2003-03       Impact factor: 3.837

8.  Materials and methods for cleaning dentures.

Authors:  E Budtz-Jørgensen
Journal:  J Prosthet Dent       Date:  1979-12       Impact factor: 3.426

9.  In vitro hardness, water sorption, and resin solubility of laboratory-processed and autopolymerized long-term resilient denture liners over one year of water storage.

Authors:  Gregory R Parr; Frederick A Rueggeberg
Journal:  J Prosthet Dent       Date:  2002-08       Impact factor: 3.426

10.  Use of tissue conditioners and resilient liners.

Authors:  J B Gonzalez
Journal:  Dent Clin North Am       Date:  1977-04
View more
  3 in total

Review 1.  [Are denture cleansing tablets suitable for daily use?].

Authors:  F H Blankenstein; I Peroz
Journal:  Z Gerontol Geriatr       Date:  2011-06       Impact factor: 1.281

2.  Denture cleanliness and hygiene: an overview.

Authors:  Petros Mylonas; Paul Milward; Robert McAndrew
Journal:  Br Dent J       Date:  2022-07-08       Impact factor: 2.727

3.  The Influence of Decontamination Procedures on the Surface of Two Polymeric Liners Used in Prosthodontics.

Authors:  Katarzyna Mańka-Malara; Maciej Trzaskowski; Dominika Gawlak
Journal:  Polymers (Basel)       Date:  2021-12-11       Impact factor: 4.329