| Literature DB >> 19089198 |
Patrícia Moreira de Freitas1, Débora Soares-Geraldo, Ana Cristina Biella-Silva, Amanda Verna Silva, Bruno Lopes da Silveira, Carlos de Paula Eduardo.
Abstract
Studies have shown the cariostatic effect of Er,Cr:YSGG (2.78 mm) laser irradiation on human enamel and have suggested its use on caries prevention. However there are still no reports on the intrapulpal temperature increase during enamel irradiation using parameters for caries prevention. The aim of this in vitro study was to evaluate the temperature variation in the pulp chamber during human enamel irradiation with Er,Cr:YSGG laser at different energy densities. Fifteen enamel blocks obtained from third molars (3 x 3 x 3 mm) were randomly assigned to 3 groups (n=5): G1--Er,Cr:YSGG laser 0.25 W, 20 Hz, 2.84 J/cm(2), G2--Er,Cr:YSGG laser 0.50 W, 20 Hz, 5.68 J/cm(2), G3--Er,Cr:YSGG laser 0.75 W, 20 Hz, 8.52 J/cm(2). During enamel irradiation, two thermocouples were fixed in the inner surface of the specimens and a thermal conducting paste was used. One-way ANOVA did not show statistically significant difference among the experimental groups (alpha=0.05). There was intrapulpal temperature variation < or = 0.1 degrees C for all irradiation parameters. In conclusion, under the tested conditions, the use of Er,Cr:YSGG laser with parameters set for caries prevention lead to an acceptable temperature increase in the pulp chamber.Entities:
Mesh:
Year: 2008 PMID: 19089198 PMCID: PMC4327626 DOI: 10.1590/s1678-77572008000200003
Source DB: PubMed Journal: J Appl Oral Sci ISSN: 1678-7757 Impact factor: 2.698
FIGURE 1Schematic diagram of the study design A and B. Third human molars were collected and had their crowns removed close to the cementoenamel junction and sectioned longitudinally. C. Fifteen 15 enamel blocks (3 x 3 mm) were obtained and further fixed individually on an acrylic device; D. Laser irradiation was carried out on the specimen of each experimental group and temperature variation (ΔT) measurements were performed during laser irradiation
Experimental groups
| Group | Power (W) | Energy Density (J/cm2) | Repetition rate (Hz) |
|---|---|---|---|
| G1 (n=5) | 0.25 | 2.8 | 20 |
| G2 (n=5) | 0.50 | 5.7 | 20 |
| G3 (n=5) | 0.75 | 8.5 | 20 |
Results of one-way analysis of variance (α=5%) for comparison of the mean values of temperature variation recorded in the experimental groups
| Among the experimental groups Experimental Groups | F 1.799 Experimental Groups | Significance 0.178 Statistical Significance (p) |
|---|---|---|
| 1 | 2 | 0.211 |
| 3 | 0.275 | |
| 2 | 1 | 0.211 |
| 3 | 0.986 | |
| 3 | 1 | 0.275 |
| 2 | 0.986 |
FIGURE 2Mean values for the temperature variation during enamel laser irradiation using caries preventive parameters