| Literature DB >> 19087350 |
Stefaan Nijs1, An Sermon, Paul Broos.
Abstract
BACKGROUND: Proximal humerus fractures are one of the most common fractures. Intramedullary locked nailing is becoming a popular alternative treatment, especially for easier fracture patterns. Although axillary nerve injury has been reported, no study has compared the safety of the proximal locking options relative to the axillary nerve and the ascending branch of the anterior circumflex artery.Entities:
Year: 2008 PMID: 19087350 PMCID: PMC2636765 DOI: 10.1186/1754-9493-2-33
Source DB: PubMed Journal: Patient Saf Surg ISSN: 1754-9493
Figure 1Evaluation of the risk of iatrogenic injury to the axillary nerve using the Sirus Nail (Zimmer. The minimal distance between the oblique inferior proximal locking bolt of the Sirus nail (Zimmer®) and the anterior branch of the axillary nerve puts this structure at risk during drilling and blunt insertion of the bolt.
Figure 2Anatomic preparation of the proximal humerus after insertion of a T2 Nail (Stryker. In the T2 Nail the limited distance between the inferior oblique proximal bolt and the anterior branch of the axillary nerve can put the nerve at risk during insertion of the bolt.
Characterstics of the six different implants used in this study.
| Synthes PHN | 6° | 150 mm | 18 mm | 90° |
| Synthes EPHN | 6° | 150 mm | 20 mm | 90° |
| Targon PH | 0° | 150 mm | 26 mm | 90° |
| Stryker T2-PHN | 6° | 150 mm | 30 mm | 105° |
| S&N Trigen PH | 0° | 160 mm | 32 mm | 90° |
| Zimmer Sirus | 9° | 120 mm | 37 mm | 120° |
Average distance of the (closest) locking bolt to the axillary nerve and the ascending branch of the anterior circumflex artery.
| Synthes PHN | 7.48 mm | 3.5 | 19.2 mm | 6.0 |
| Synthes EPHN | 6.4 mm | 5,1 | 14.2 mm | 3.0 |
| Targon PH | 13.6 mm | 8,1 | 5.8 mm | 3.1 |
| Stryker T2-PHN | 2.7 mm | 1,1 | 8 mm | 4.3 |
| S&N Trigen PH | 10 mm | 2,7 | 8.1 mm | 2.1 |
| Zimmer Sirus | 1 mm | 2 | 14.1 mm | 3.9 |
Mean distance of the (closest) locking bolt to the theoretical average position of the axillary nerve
| Synthes PHN | 8.6 mm | 3.4 |
| Synthes EPHN | 8.2 mm | 6.0 |
| Targon PH | 15.7 mm | 2.1 |
| Stryker T2-PHN | 2.7 mm | 1.9 |
| S&N Trigen PH | 6.7 mm | 5.2 |
| Zimmer Sirus | 5.0 mm | 2.3 |