Chih-Hsin Tsai1, Jia-Der Lin, Cheng-Huang Lin.
Abstract
A capillary electrophoresis Raman spectroscopy (CE-RS) method based on the stacking and sweeping modes are described. A non-fluorescent compound (malachite green, MG; crystal violet, CV) and a doubled Nd:YAG laser (532 nm, 300 mW) were selected as the model compound and light source, respectively. In order to carry out a quantitative and analysis of MG, a monochromator was used to collect the specific Raman line at 1616 cm(-1) (the N-phi and C-C stretching, corresponding to 582 nm when the wavelength of the exciting source is 532 nm). The limit of detection (LOD) for MG was 1.6 x 10(-5) and 1.1 x 10(-5)M, respectively, based on the CZE and MEKC modes. This could be improved to 3.4 x 10(-7) and 5.3 x 10(-9)M, respectively, when the stacking and sweeping modes were applied. The method was also extended to the determination of MG in an actual sample.Entities:
Year: 2006 PMID: 19071626 DOI: 10.1016/j.talanta.2006.10.037
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057